Development and validation of a novel qualitative test for plasma fibrinogen utilizing clot waveform analysis.
Ontology highlight
ABSTRACT: Plasma fibrinogen is commonly examined by Clauss fibrinogen assay, which cannot distinguish between quantitative and qualitative fibrinogen anomalies. However, our previously reported Clauss fibrinogen assay utilizing clot waveform analysis (Clauss-CWA) provides additional information that contributes to the classification of fibrinogen anomalies. In this study, we adopted the Clauss-CWA method for an autoanalyzer to automatically measure the antigenic estimate (eAg) of fibrinogen in addition to the functional amount (Ac), and to thus provide the Ac/eAg ratio as a qualitative indicator. Performance was validated by receiver operating characteristics (ROC) and precision recall (PR) curve analyses using a patient cohort, consisting of a training cohort (n = 519) and a validation cohort (n = 523), both of which contained cases of congenital (hypo)dysfibrinogenemia as qualitative defects. We obtained an optimal cutoff of 0.65 for Ac/eAg by ROC curve analysis of the training cohort, offering superior sensitivity (> 0.9661) and specificity (1.000). This cutoff was validated in the validation cohort, providing positive predictive value > 0.933 and negative predictive value > 0.998. PR curve analysis also showed that Clauss-CWA provided excellent performance for detecting qualitative fibrinogen anomalies. The Clauss-CWA method may represent a useful approach for detecting qualitative fibrinogen abnormalities in routine laboratory testing.