Performance of bioelectrical impedance analysis compared to dual X-ray absorptiometry (DXA) in Veterans with COPD.


ABSTRACT: We examined the performance of a commercially-available handheld bioimpedance (BIA) device relative to dual X-ray absorptiometry (DXA) to assess body composition differences among Veterans with chronic obstructive pulmonary disease (COPD). Body composition was measured using DXA and BIA (Omron HBF-306C) at a single time point. Correlations between BIA- and DXA-assessed percent fat, fat mass, and fat-free mass were analyzed using Spearman (ρ) and Lin Concordance Correlation Coefficients (ρc). Mean differences in fat mass were visualized using Bland-Altman plots. Subgroup analyses by obesity status (BMI < 30 versus ≥ 30) were performed. Among 50 participants (96% male; mean age: 69.5 ± 6.0 years), BIA-assessed fat mass was strongly correlated (ρ = 0.94) and demonstrate excellent concordance (ρc = 0.95, [95%CI: 0.93-0.98]) with DXA, with a mean difference of 2.7 ± 3.2 kg between BIA and DXA. Although Spearman correlations between BIA- and DXA-assessed percent fat and fat-free mass were strong (ρ = 0.8 and 0.91, respectively), concordance values were only moderate (ρc = 0.67 and 0.74, respectively). Significantly stronger correlations were observed for obese relative to non-obese subjects for total percent fat (ρobese = 0.85 versus ρnon-obese = 0.5) and fat mass (ρobese = 0.96 versus ρnon-obese = 0.84). A handheld BIA device demonstrated high concordance with DXA for fat mass and moderate concordance for total percent fat and fat-free mass.ClinicalTrials.gov: NCT02099799.

SUBMITTER: Cruz Rivera PN 

PROVIDER: S-EPMC8816927 | biostudies-literature | 2022 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance of bioelectrical impedance analysis compared to dual X-ray absorptiometry (DXA) in Veterans with COPD.

Cruz Rivera Paola N PN   Goldstein Rebekah L RL   Polak Madeline M   Lazzari Antonio A AA   Moy Marilyn L ML   Wan Emily S ES  

Scientific reports 20220204 1


We examined the performance of a commercially-available handheld bioimpedance (BIA) device relative to dual X-ray absorptiometry (DXA) to assess body composition differences among Veterans with chronic obstructive pulmonary disease (COPD). Body composition was measured using DXA and BIA (Omron HBF-306C) at a single time point. Correlations between BIA- and DXA-assessed percent fat, fat mass, and fat-free mass were analyzed using Spearman (ρ) and Lin Concordance Correlation Coefficients (ρ<sub>c<  ...[more]

Similar Datasets

| S-EPMC6433552 | biostudies-literature
| S-EPMC9077186 | biostudies-literature
| S-EPMC7263310 | biostudies-literature
| S-EPMC7004364 | biostudies-literature
| S-EPMC11366593 | biostudies-literature
| S-EPMC7177846 | biostudies-literature
| S-EPMC7230013 | biostudies-literature
| S-EPMC10928253 | biostudies-literature
| S-EPMC7751214 | biostudies-literature