Accelerated Breast Diffusion-weighted Imaging Using Multiband Sensitivity Encoding with the CAIPIRINHA Method: Clinical Experience at 3 T.
Ontology highlight
ABSTRACT: Purpose To examine the clinical value of multiband (MB) sensitivity encoding (SENSE)-accelerated diffusion-weighted imaging (DWI) for breast imaging by performing quantitative and qualitative comparisons with conventional diffusion-weighted echo-planar imaging, or conventional DWI (cDWI). Materials and Methods In this prospective study (ClinicalTrials.gov identifier NCT03607552), women with breast cancer were recruited from July 2018 to July 2019 to undergo additional MB SENSE DWI during clinical 3-T breast MRI examinations. The cDWI and MB SENSE DWI acquisitions were assessed both quantitatively and qualitatively. Regions of interest were defined for tumorous and normal tissue, and the tumor apparent diffusion coefficient (ADC), contrast-to-noise ratio (CNR), and signal index (SI) were calculated for both DWI methods. Three readers independently reviewed the two acquisitions side by side and provided relative image quality scores. Tumor ADC, CNR, and SI measures were compared between cDWI and MB SENSE DWI acquisitions by using a paired t test, and reader preferences were evaluated by using the sign test. Results The study included 38 women (median age, 48 years; range, 28-83 years). Overall agreement was good between cDWI and MB SENSE DWI tumor ADC measures (intraclass correlation coefficient, 0.87 [95% CI: 0.75, 0.94]), and no differences were evident in the ADC (median, 0.93 × 10-3 mm2/sec vs 0.87 ×10-3 mm2/sec; P = .50), CNR (2.2 vs 2.3; P = .17), or SI (9.2 vs 9.2; P = .23) measurements. The image quality of cDWI and MB SENSE DWI acquisitions were considered equal for 51% of images (58 of 114), whereas MB SENSE DWI was preferred more often than cDWI (37% [42 of 114] vs 12% [14 of 114]; P < .001). The preference for MB SENSE DWI was most often attributed to better fat suppression. Conclusion MB SENSE can be used to accelerate breast DWI acquisition times without compromising the image quality or the fidelity of quantitative ADC measurements. Keywords: MR-Diffusion-weighted Imaging, Breast, Comparative Studies, Technology Assessment Clinical trial registration no. NCT03607552 © RSNA, 2022.
SUBMITTER: Biswas D
PROVIDER: S-EPMC8830507 | biostudies-literature |
REPOSITORIES: biostudies-literature
ACCESS DATA