Development and Validation of a Reliable UHPLC-MS/MS Method for Simultaneous Quantification of Macrocyclic Lactones in Bovine Plasma.


ABSTRACT: A fast, accurate and reliable ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method was developed for simultaneous quantification of ivermectin (IVER), doramectin (DORA), and moxidectin (MOXI) in bovine plasma. A priority for sample preparation was the eradication of possible infectious diseases to avoid travel restrictions. The sample preparation was based on protein precipitation using 1% formic acid in acetonitrile, followed by Ostro® 96-well plate pass-through sample clean-up. The simple and straightforward procedure, along with the short analysis time, makes the current method unique and suitable for a large set of sample analyses per day for PK studies. Chromatographic separation was performed using an Acquity UPLC HSS-T3 column, with 0.01% acetic acid in water and methanol, on an Acquity H-Class ultra-high performance liquid chromatograph (UHPLC) system. The MS/MS instrument was a Xevo TQ-S® mass spectrometer, operating in the positive electrospray ionization mode and two multiple reaction monitoring (MRM) transitions were monitored per component. The MRM transitions of m/z 897.50 > 753.4 for IVER, m/z 921.70 > 777.40 for DORA and m/z 640.40 > 123.10 for MOXI were used for quantification. The method validation was performed using matrix-matched calibration curves in a concentration range of 1 to 500 ng/mL. Calibration curves fitted a quadratic regression model with 1/x2 weighting (r ≥ 0.998 and GoF ≤ 4.85%). Limits of quantification (LOQ) values of 1 ng/mL were obtained for all the analytes, while the limits of detection (LOD) were 0.02 ng/mL for IVER, 0.03 ng/mL for DORA, and 0.58 ng/mL for MOXI. The results of within-day (RSD < 6.50%) and between-day (RSD < 8.10%) precision and accuracies fell within acceptance ranges. No carry-over and no peak were detected in the UHPLC-MS/MS chromatogram of blank samples showing good specificity of the method. The applicability of the developed method was proved by an analysis of the field PK samples.

SUBMITTER: Zeleke G 

PROVIDER: S-EPMC8838099 | biostudies-literature | 2022 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Development and Validation of a Reliable UHPLC-MS/MS Method for Simultaneous Quantification of Macrocyclic Lactones in Bovine Plasma.

Zeleke Gemechu G   De Baere Siegrid S   Suleman Sultan S   Devreese Mathias M  

Molecules (Basel, Switzerland) 20220201 3


A fast, accurate and reliable ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method was developed for simultaneous quantification of ivermectin (IVER), doramectin (DORA), and moxidectin (MOXI) in bovine plasma. A priority for sample preparation was the eradication of possible infectious diseases to avoid travel restrictions. The sample preparation was based on protein precipitation using 1% formic acid in acetonitrile, followed by Ostro<sup>®</sup> 96-well pl  ...[more]

Similar Datasets

| S-EPMC8153023 | biostudies-literature
| S-EPMC6749594 | biostudies-literature
| S-EPMC9293867 | biostudies-literature
| S-EPMC8891122 | biostudies-literature
| S-EPMC9399508 | biostudies-literature
| S-EPMC8953653 | biostudies-literature
| S-EPMC8541045 | biostudies-literature
| S-EPMC6559858 | biostudies-literature
| S-EPMC6446716 | biostudies-literature
| S-EPMC8143330 | biostudies-literature