Characterizing Fixational Eye Motion Variance Over Time as Recorded by the Tracking Scanning Laser Ophthalmoscope.


ABSTRACT:

Purpose

The purpose of this study was to characterize the benign biological variance of fixational microsaccades in a control population using a tracking scanning laser ophthalmoscope (TSLO), accounting for machine accuracy and precision, to determine ideal testing conditions to detect pathologic change in fixational eye motion (FEM).

Methods

We quantified the accuracy and precision of the TSLO, analyzing measurements made by three operators on a model eye. Repeated, 10-second retinal motion traces were then recorded in 17 controls, 3 times a day (morning, afternoon, and evening), on 3 separate days. Microsaccade metrics (MMs) of frequency, average amplitude, peak velocity, and peak acceleration were extracted. Trace to trace, interday, and intraday variability were calculated across all subjects.

Results

Intra-operator and machine variation contributed minimally to total variation, with only 0.007% and 0.14% contribution for frequency and amplitude respectively. Bias was detected, with lower accuracy for higher amplitudes. Participants had an average (SD) microsaccade frequency of 0.84 Hz (0.52 Hz), amplitude of 0.32 degrees (0.11 degrees), peak velocity of 43.68 degrees/s (14.02 degrees/s), and peak acceleration of 13,920.04 degrees/s2 (4,186.84 degrees/s2). The first trace recorded within a session significantly differed from the second two in both microsaccade acceleration and velocity (P < 0.05), and frequency was 0.098 Hz higher in the evenings (P < 0.05). There was no MM difference between days and no evidence of a session-level learning effect (P > 0.05).

Conclusions

The TSLO is both accurate and precise. However, biological inter- and intra-individual variance is present. Trace to trace variability and time of day should be accounted for to optimize detection of pathologic change.

SUBMITTER: Condor Montes SY 

PROVIDER: S-EPMC8883154 | biostudies-literature | 2022 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Characterizing Fixational Eye Motion Variance Over Time as Recorded by the Tracking Scanning Laser Ophthalmoscope.

Condor Montes Shivany Y SY   Bennett Daniel D   Bensinger Ethan E   Rani Lakshmisahithi L   Sherkat Younes Y   Zhao Chao C   Helft Zachary Z   Roorda Austin A   Green Ari J AJ   Sheehy Christy K CK  

Translational vision science & technology 20220201 2


<h4>Purpose</h4>The purpose of this study was to characterize the benign biological variance of fixational microsaccades in a control population using a tracking scanning laser ophthalmoscope (TSLO), accounting for machine accuracy and precision, to determine ideal testing conditions to detect pathologic change in fixational eye motion (FEM).<h4>Methods</h4>We quantified the accuracy and precision of the TSLO, analyzing measurements made by three operators on a model eye. Repeated, 10-second ret  ...[more]

Similar Datasets

| S-EPMC5378343 | biostudies-literature
| S-EPMC4505698 | biostudies-other
| S-EPMC3469984 | biostudies-other
| S-EPMC3348404 | biostudies-literature
| S-EPMC4740427 | biostudies-literature
| S-EPMC8759795 | biostudies-literature
| S-EPMC7316009 | biostudies-literature
| S-EPMC3250263 | biostudies-literature
| S-EPMC7278210 | biostudies-literature
| S-EPMC3005164 | biostudies-other