Diagnostic accuracy of SARS-CoV-2 saliva antigen testing in a real-life clinical setting.


ABSTRACT:

Background

SARS-CoV-2 antigen tests with saliva facilitate examination in settings that lack trained personnel. However, little is known about the diagnostic accuracy in real-life clinical settings. Therefore, we studied the diagnostic accuracy of a saliva antigen test in diagnosing SARS-CoV-2 infection in a primary/secondary care testing facility.

Methods

Individuals who presented at a COVID-19 testing facility affiliated with a Swiss university hospital were prospectively recruited (n=377). Saliva specimen was obtained, and the PCL Inc. COVID19 Gold antigen test was conducted in parallel with 2 real-time polymerase chain reaction (RT-PCR) assays from a nasopharyngeal swab.

Results

RT-PCR results were positive in 53 individuals, corresponding to a prevalence of 14.1% (missing material in 1 individual). The PCL saliva antigen test was positive in 22 individuals (5.8%) and negative in 354 (93.9%). The sensitivity of the saliva antigen test was 30.2% (95% confidence interval 18.3, 44.3), both overall and in symptomatic individuals. The specificity was 98.1% (96.0, 99.3).

Conclusions

The diagnostic accuracy of a SARS-CoV-2 saliva antigen test in a primary/secondary care testing facility was remarkably lower than that reported in the manufacturer's specifications.

SUBMITTER: Jegerlehner S 

PROVIDER: S-EPMC8964446 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC8321236 | biostudies-literature
| S-EPMC8133768 | biostudies-literature
| S-EPMC7943555 | biostudies-literature
| S-EPMC8163726 | biostudies-literature
| S-EPMC9590385 | biostudies-literature
| S-EPMC10010503 | biostudies-literature
| S-EPMC8457903 | biostudies-literature
| S-EPMC9016156 | biostudies-literature
| S-EPMC6102559 | biostudies-literature
| S-EPMC7553188 | biostudies-literature