Improvement of osseointegration of Ti-6Al-4V ELI alloy orthodontic mini-screws through anodization, cyclic pre-calcification, and heat treatments.


ABSTRACT:

Background

Mini-screws are widely used as temporary anchorages in orthodontic treatment, but have the disadvantage of showing a high failure rate of about 10%. Therefore, orthodontic mini-screws should have high biocompatibility and retention. Previous studies have demonstrated that the retention of mini-screws can be improved by imparting bioactivity to the surface. The method for imparting bioactivity proposed in this paper is to sequentially perform anodization, periodic pre-calcification, and heat treatments with a Ti-6Al-4V ELI alloy mini-screw.

Materials and methods

A TiO2 nanotube-structured layer was formed on the surface of the Ti-6Al-4V ELI alloy mini-screw through anodization in which a voltage of 20 V was applied to a glycerol solution containing 20 wt% H2O and 1.4 wt% NH4F for 60 min. Fine granular calcium phosphate precipitates of HA and octacalcium phosphate were generated as clusters on the surface through the cyclic pre-calcification and heat treatments. The cyclic pre-calcification treatment is a process of immersion in a 0.05 M NaH2PO4 solution and a saturated Ca(OH)2 solution at 90 °C for 1 min each.

Results

It was confirmed that the densely structured protrusions were precipitated, and Ca and P concentrations, which bind and concentrate endogenous bone morphogenetic proteins, increased on the surface after simulated body fluid (SBF) immersion test. In addition, the removal torque of the mini-screw fixed into rabbit tibias for 4 weeks was measured to be 8.70 ± 2.60 N cm.

Conclusions

A noteworthy point in this paper is that the Ca and P concentrations, which provide a scaffold suitable for endogenous bone formation, further increased over time after SBF immersion of the APH group specimens. The other point is that our mini-screws have a significantly higher removal torque compared to untreated mini-screws. These results represent that the mini-screw proposed in this paper can be used as a mini-screw for orthodontics.

SUBMITTER: Im C 

PROVIDER: S-EPMC8977256 | biostudies-literature | 2022 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Improvement of osseointegration of Ti-6Al-4V ELI alloy orthodontic mini-screws through anodization, cyclic pre-calcification, and heat treatments.

Im Changkyun C   Park Je-Hyeok JH   Jeon Young-Mi YM   Kim Jong-Ghee JG   Jang Yong-Seok YS   Lee Min-Ho MH   Jeon Woo-Yong WY   Kim Jun-Min JM   Bae Tae-Sung TS  

Progress in orthodontics 20220404 1


<h4>Background</h4>Mini-screws are widely used as temporary anchorages in orthodontic treatment, but have the disadvantage of showing a high failure rate of about 10%. Therefore, orthodontic mini-screws should have high biocompatibility and retention. Previous studies have demonstrated that the retention of mini-screws can be improved by imparting bioactivity to the surface. The method for imparting bioactivity proposed in this paper is to sequentially perform anodization, periodic pre-calcifica  ...[more]

Similar Datasets

| S-EPMC9547804 | biostudies-literature
| S-EPMC5458061 | biostudies-literature
| S-EPMC4761652 | biostudies-literature
| S-EPMC5820288 | biostudies-literature
| S-EPMC6170433 | biostudies-literature
| S-EPMC7817818 | biostudies-literature
| S-EPMC8400223 | biostudies-literature
| S-EPMC10995190 | biostudies-literature
| S-EPMC7504553 | biostudies-literature
| S-EPMC7289882 | biostudies-literature