Accuracy of atrial fibrillation detection by an insertable cardiac monitor in patients undergoing catheter ablation: Results of the BioVAD study.


ABSTRACT:

Background

Insertable cardiac monitors (ICMs) are increasingly used to evaluate the atrial fibrillation (AF) burden after catheter ablation of AF. BioMonitor III (BM3) is an ICM with a long sensing vector, which enhances sensing capabilities. The AF detection algorithm of the BM3 is based on R-R interval variability.

Objective

To evaluate the performance of the AF detection algorithm of BM3 in patients before and after catheter ablation of AF using simultaneous Holter recordings.

Methods

In this prospective study, we enrolled patients scheduled for catheter ablation of paroxysmal or persistent AF. After BM3 implantation, patients had a 4 days Holter registration before and 3 months after ablation. All true AF episodes ≥2 min on the Holter were annotated and matched with BM3 detected AF detections.

Results

Thirty-one patients were enrolled (mean age 60 ± 8, 74% male, 68% paroxysmal AF). Fifty-six Holter registrations were performed in 30 patients. Twelve patients demonstrated at least one true AF episode with a total AF duration of 570 h. The AF burden accuracy of BM3 before catheter ablation was 99.6%, with a duration sensitivity of 98.6% and a duration specificity of 99.9%. The AF burden accuracy of BM3 after catheter ablation was 99.8%, with a duration sensitivity of 90.2% and a duration specificity of 99.9%. Overall, the AF burden detected on the Holter and BM3 demonstrated a high Pearson correlation coefficient of 0.996.

Conclusion

BM3 accurately detects AF burden in patients before and after catheter ablation of AF.

SUBMITTER: Assaf A 

PROVIDER: S-EPMC9107077 | biostudies-literature | 2022 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy of atrial fibrillation detection by an insertable cardiac monitor in patients undergoing catheter ablation: Results of the BioVAD study.

Assaf Amira A   Theuns Dominic A M J DAMJ   Sakhi Rafi R   Bhagwandien Rohit E RE   Szili-Torok Tamas T   Yap Sing-Chien SC  

Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc 20220428 3


<h4>Background</h4>Insertable cardiac monitors (ICMs) are increasingly used to evaluate the atrial fibrillation (AF) burden after catheter ablation of AF. BioMonitor III (BM3) is an ICM with a long sensing vector, which enhances sensing capabilities. The AF detection algorithm of the BM3 is based on R-R interval variability.<h4>Objective</h4>To evaluate the performance of the AF detection algorithm of BM3 in patients before and after catheter ablation of AF using simultaneous Holter recordings.<  ...[more]

Similar Datasets

| S-EPMC7237079 | biostudies-literature
| S-EPMC3500836 | biostudies-literature
| S-EPMC9761217 | biostudies-literature
| S-EPMC4090097 | biostudies-other
| S-EPMC10382095 | biostudies-literature
2025-02-14 | GSE289284 | GEO
| S-EPMC11199813 | biostudies-literature
| S-EPMC6504925 | biostudies-literature
| S-EPMC6393682 | biostudies-literature
| S-EPMC1860761 | biostudies-literature