Design and Characterization of an RF Applicator for In Vitro Tests of Electromagnetic Hyperthermia.


ABSTRACT: The evaluation of the biological effects of therapeutic hyperthermia in oncology and the precise quantification of thermal dose, when heating is coupled with radiotherapy or chemotherapy, are active fields of research. The reliable measurement of hyperthermia effects on cells and tissues requires a strong control of the delivered power and of the induced temperature rise. To this aim, we have developed a radiofrequency (RF) electromagnetic applicator operating at 434 MHz, specifically engineered for in vitro tests on 3D cell cultures. The applicator has been designed with the aid of an extensive modelling analysis, which combines electromagnetic and thermal simulations. The heating performance of the built prototype has been validated by means of temperature measurements carried out on tissue-mimicking phantoms and aimed at monitoring both spatial and temporal temperature variations. The experimental results demonstrate the capability of the RF applicator to produce a well-focused heating, with the possibility of modulating the duration of the heating transient and controlling the temperature rise in a specific target region, by simply tuning the effectively supplied power.

SUBMITTER: Ferrero R 

PROVIDER: S-EPMC9148047 | biostudies-literature | 2022 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Design and Characterization of an RF Applicator for In Vitro Tests of Electromagnetic Hyperthermia.

Ferrero Riccardo R   Androulakis Ioannis I   Martino Luca L   Nadar Robin R   van Rhoon Gerard C GC   Manzin Alessandra A  

Sensors (Basel, Switzerland) 20220510 10


The evaluation of the biological effects of therapeutic hyperthermia in oncology and the precise quantification of thermal dose, when heating is coupled with radiotherapy or chemotherapy, are active fields of research. The reliable measurement of hyperthermia effects on cells and tissues requires a strong control of the delivered power and of the induced temperature rise. To this aim, we have developed a radiofrequency (RF) electromagnetic applicator operating at 434 MHz, specifically engineered  ...[more]

Similar Datasets

| S-EPMC7347919 | biostudies-literature
| S-EPMC6721488 | biostudies-literature
| S-EPMC3866607 | biostudies-other
| S-EPMC3701653 | biostudies-literature
| S-EPMC10340829 | biostudies-literature
| S-EPMC8657004 | biostudies-literature
| S-EPMC5697703 | biostudies-literature
| S-EPMC5356646 | biostudies-literature
| S-EPMC10005510 | biostudies-literature
| S-EPMC6334175 | biostudies-literature