ABSTRACT: Purpose
The purpose of this study was to evaluate three different methods of attachment of continuous loop suspensory cortical preparation of all soft tissue central quad tendon grafts compared to a bone block control for anterior cruciate ligament reconstruction on construct displacement and load to failure.Methods
Thirty-two cadaveric central quadriceps tendon (CQT) specimens were harvested, using three clinical techniques for graft fixation: cortical button alone (BTB EB), BTB cortical button with rip-stop suture (BTB RS), and continuous loop cortical button (BTB CL). A control group was also included that consisted of a bone block secured within testing clamps (BTB CON). Specimens were preloaded to 150 N. Tendons were then cyclically loaded between 50 N and 250 N for 1,