Unknown

Dataset Information

0

Paper Biosensor for the Detection of NT-proBNP Using Silver Nanodisks as Electrochemical Labels.


ABSTRACT: We report on the use of silver nanodisks (AgNDs), having a diameter of 50 ± 8 nm and a thickness of 8 ± 2 nm, as electrochemical labels for the detection of a model metalloimmunoassay for the heart failure biomarker NT-proBNP. The detection method is based on an electrochemically activated galvanic exchange (GE) followed by the detection of Ag using anodic stripping voltammetry (ASV). The AgNDs labels are superior to Ag nanocubes and Ag nanospheres in terms of the dynamic range for both the model and NT-proBNP metalloimmunoassays. The linear dynamic range for the model composite is 1.5 to 30.0 pM AgNDs. When AgND labels are used for the NT-proBNP assay, the dynamic range is 0.03-4.0 nM NT-proBNP. The latter range fully overlaps the risk stratification range for heart failure from 53 pM to 590 pM. The performance improvement of the AgNDs is a result of the specific GE mechanism for nanodisks. Specifically, GE is complete across the face of the AgNDs, leaving behind an incompletely exchanged ring structure composed of both Ag and Au.

SUBMITTER: Peng Y 

PROVIDER: S-EPMC9268099 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC7103555 | biostudies-literature
| S-EPMC9178520 | biostudies-literature
| S-EPMC10888530 | biostudies-literature
| S-EPMC7180619 | biostudies-literature
| S-EPMC10620918 | biostudies-literature
| S-EPMC8005456 | biostudies-literature
| S-EPMC8740487 | biostudies-literature
| S-EPMC10377000 | biostudies-literature
| S-EPMC5336011 | biostudies-literature
| S-EPMC7746088 | biostudies-literature