Given the increasing use of photo-activated resins in dentistry, dentists and researchers need a user-friendly dental radiometer to measure the power output from dental light-curing units (LCUs).Objective
Our goal was to measure the accuracy of two brands of dental radiometers in reporting the power (mW) from twelve brands of contemporary LCUs compared to a 'gold standard' (GS) reference value obtained from an integrating sphere attached to a fiberoptic spectroradiometer.Methods
The power output was measured from two units of 12 brands of LCUs, five times on the ''GS" system, five times on two Bluephase Meter II dental radiometers, and five times on two Mini Gig hand-held spectroradiometers. The emission spectrum was also recorded using the 'GS' integrating sphere. The power values reported by each meter were subjected to t-tests to compare the two examples of each LCU, and 3-way ANOVA followed by Bonferroni's post-hoc tests. Regression analyses were also performed to determine the relationship between the data from the hand-held radiometers and the 'GS' integrating sphere.Results
There was a large difference in the power values (mW) and the emission spectra from the 12 brands of LCUs on their standard-settings (p<0.001). Except for one LCU (Dental Spark @ 15.1%), the differences between the two LCUs of the same brand were less than 5.3% when measured using the 'GS' integrating sphere. Regression analyses showed a highly significant agreement between the power values reported from the two brands of radiometers and the 'GS' integrating sphere (R2 > 98%).Conclusion
We concluded that the power values reported from both brands of dental radiometers we tested were accurate, provided that the light source did not emit wavelengths of light that were beyond the radiometer's detection limit.