Power output from 12 brands of contemporary LED light-curing units measured using 2 brands of radiometers.


ABSTRACT:

Background

Given the increasing use of photo-activated resins in dentistry, dentists and researchers need a user-friendly dental radiometer to measure the power output from dental light-curing units (LCUs).

Objective

Our goal was to measure the accuracy of two brands of dental radiometers in reporting the power (mW) from twelve brands of contemporary LCUs compared to a 'gold standard' (GS) reference value obtained from an integrating sphere attached to a fiberoptic spectroradiometer.

Methods

The power output was measured from two units of 12 brands of LCUs, five times on the ''GS" system, five times on two Bluephase Meter II dental radiometers, and five times on two Mini Gig hand-held spectroradiometers. The emission spectrum was also recorded using the 'GS' integrati

SUBMITTER: Maucoski C 

PROVIDER: S-EPMC9269870 | biostudies-literature | 2022