Reduced Chest Computed Tomography Scan Length for Patients Positive for Coronavirus Disease 2019: Dose Reduction and Impact on Diagnostic Utility.


ABSTRACT:

Methods

This study used the Personalized Rapid Estimation of Dose in CT (PREDICT) tool to estimate patient-specific organ doses from CT image data. The PREDICT is a research tool that combines a linear Boltzmann transport equation solver for radiation dose map generation with deep learning algorithms for organ contouring. Computed tomography images from 74 subjects in the Medical Imaging Data Resource Center-RSNA International COVID-19 Open Radiology Database data set (chest CT of adult patients positive for COVID-19), which included expert annotations including "infectious opacities," were analyzed. First, the full z-scan length of the CT image data set was evaluated. Next, the z-scan length was reduced from the left hemidiaphragm to the top of the aortic arch. Generic dose reduction based on dose length product (DLP) and patient-specific organ dose reductions were calculated. The percentage of infectious opacities excluded from the reduced z-scan length was used to quantify the effect on diagnostic utility.

Results

Generic dose reduction, based on DLP, was 69%. The organ dose reduction ranged from approximately equal to 18% (breasts) to approximately equal to 64% (bone surface and bone marrow). On average, 12.4% of the infectious opacities were not included in the reduced z-coverage, per patient, of which 5.1% were above the top of the arch and 7.5% below the left hemidiaphragm.

Conclusions

Limiting z-scan length of chest CTs reduced radiation dose without significantly compromising diagnostic utility in COVID-19 patients. The PREDICT demonstrated that patient-specific organ dose reductions varied from generic dose reduction based on DLP.

SUBMITTER: Principi S 

PROVIDER: S-EPMC9296570 | biostudies-literature | 2022 Jul-Aug 01

REPOSITORIES: biostudies-literature

altmetric image

Publications

Reduced Chest Computed Tomography Scan Length for Patients Positive for Coronavirus Disease 2019: Dose Reduction and Impact on Diagnostic Utility.

Principi Sara S   O'Connor Stacy S   Frank Luba L   Schmidt Taly Gilat TG  

Journal of computer assisted tomography 20220408 4


<h4>Methods</h4>This study used the Personalized Rapid Estimation of Dose in CT (PREDICT) tool to estimate patient-specific organ doses from CT image data. The PREDICT is a research tool that combines a linear Boltzmann transport equation solver for radiation dose map generation with deep learning algorithms for organ contouring. Computed tomography images from 74 subjects in the Medical Imaging Data Resource Center-RSNA International COVID-19 Open Radiology Database data set (chest CT of adult  ...[more]

Similar Datasets

| S-EPMC7567660 | biostudies-literature
| S-EPMC7447615 | biostudies-literature
| S-EPMC8811401 | biostudies-literature
| S-EPMC8353969 | biostudies-literature
| S-EPMC7290647 | biostudies-literature
| S-EPMC7248616 | biostudies-literature
| S-EPMC8588981 | biostudies-literature
| S-EPMC7191895 | biostudies-literature
| S-EPMC7549426 | biostudies-literature
| S-EPMC7908963 | biostudies-literature