Comparing the clinical utility of single-shot, readout-segmented and zoomit echo-planar imaging in diffusion-weighted imaging of the kidney at 3 T.
Ontology highlight
ABSTRACT: We compared the clinical utility of single-shot echo-planar imaging (SS-EPI) using different breathing schemes, readout-segmented EPI and zoomit EPI in the repeatability of apparent diffusion coefficient (ADC) measurements, cortico-medullary contrast to noise ratio (c-mCNR) and image quality. In this institutional review board-approved prospective study, some common clinically applicable diffusion-weighted imaging (b = 50, 400, 800 s/mm2) of kidney on 3.0 T MRI were performed on 22 volunteers using SS-EPI with breath-hold diffusion-weighted imaging (BH-DWI), free-breathing (FB-DWI), navigator-triggered (NT-DWI) and respiratory-triggered (RT-DWI), readout-segmented DWI (RS-DWI), and Zoomit DWI (Z-DWI). ADC and c-mCNR were measured in 12 anatomic locations (the upper, middle, and