ABSTRACT: Purpose
To evaluate fundus examination accuracy of medical students when using an unmodified iPhone X or a direct ophthalmoscope in comparison to a staff ophthalmologist's retinal examination.Methods
In this prospective comparative analysis, patients underwent dilated fundus examination by novice medical trainees using either an unmodified iPhone X or standard direct ophthalmoscope. The primary outcome was the mean difference and degree of agreement in cup-to-disc ratio between student examination and the staff ophthalmologist's cup-to-disc observation.Results
A total of 18 medical students conducted 230 retinal examinations, 117 with the iPhone X and 113 with the direct ophthalmoscope. A greater proportion of students were unable to report cup-to-disc ratio using the iPhone X (81.2%) vs direct ophthalmoscope (30.1%). Student examination of cup-to-disc ratio led to a systematic bias (95% limits of agreement) of + 0.16 (-0.22 to + 0.54) and + 0.10 (-0.36 to + 0.56) with the iPhone X and direct ophthalmoscope, respectively. iPhone X and direct ophthalmoscope student observation concordance for optic disc colour (88.7 and 82.4%, respectively) and contour (68.3 and 74.2%, respectively) demonstrated low agreement with staff ophthalmologist findings. Student iPhone X observations demonstrated lower agreement with staff findings compared to direct ophthalmoscope observations for spontaneous venous pulsations (Cohen's Kappa = -0.044 vs 0.099).Conclusion
Amongst medical trainees, optic disc visualization using an unmodified iPhone X was inferior to the direct ophthalmoscope. When able to visualize the optic nerve head, there was no significant difference in reported cup-to-disc ratio between modalities. However, both modalities demonstrated poor reliability in comparison to staff ophthalmologist findings.
SUBMITTER: Ahmed Y
PROVIDER: S-EPMC9308475 | biostudies-literature |
REPOSITORIES: biostudies-literature