Preclinical evaluation of thin convex probe endobronchial ultrasound-guided transbronchial needle aspiration for intrapulmonary lesions.


ABSTRACT:

Background

Conventional flexible bronchoscopy has not achieved the high diagnostic yield for intrapulmonary lesions as seen with image-guided transthoracic biopsy. A thin convex probe endobronchial ultrasound bronchoscope (TCP-EBUS) with a 5.9-mm tip was designed to improve peripheral access over conventional EBUS bronchoscopes to facilitate real-time sampling of intrapulmonary lesions under ultrasound guidance.

Methods

TCP-EBUS was inserted into the distal airways of ex-vivo human lungs to assess bronchial accessibility relative to clinically available bronchoscopes. The short- (≤1 h) and medium-term (≤10 d) safety of TCP-EBUS insertion and EBUS-guided transbronchial needle aspiration (TBNA) using a 25-gauge needle were evaluated physiologically and radiologically in live pigs. TCP-EBUS-guided TBNA feasibility was assessed in-vivo with pig intrapulmonary pseudo-tumors and ex-vivo with resected human lung cancer specimens.

Results

For bronchial accessibility, TCP-EBUS demonstrated greater reach than the 6.6-mm convex probe endobronchial ultrasound (CP-EBUS) in all bronchi, as well as surpassed a 5.5-mm conventional bronchoscope in 63% (131/209) and a 4.8-mm conventional bronchoscope in 27% (57/209) of assessed bronchi. The median bronchial generation and the mean diameter of bronchi TCP-EBUS reached was 4 (range, 3-7) and 3.3±0.7 mm, respectively. No major complications related to TCP-EBUS-guided TBNA in distal airways were observed in the live pigs. Scattered mucosal erythema of the bronchial walls was observed immediately after TCP-EBUS insertion; this self-resolved by day 10. TCP-EBUS could successfully reach and visualize intrapulmonary targets via ultrasound, with no difficulty in needle deployment or sampling.

Conclusions

TCP-EBUS has the potential to facilitate safe real-time transbronchial sampling of intrapulmonary lesions in the central and middle lung fields.

SUBMITTER: Ishiwata T 

PROVIDER: S-EPMC9359960 | biostudies-literature | 2022 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Preclinical evaluation of thin convex probe endobronchial ultrasound-guided transbronchial needle aspiration for intrapulmonary lesions.

Ishiwata Tsukasa T   Inage Terunaga T   Gregor Alexander A   Motooka Yamato Y   Chan Harley H L HHL   Bernards Nicholas N   Aragaki Masato M   Chen Zhenchian Z   Ujiie Hideki H   Kinoshita Tomonari T   Effat Andrew A   Yasufuku Kazuhiro K  

Translational lung cancer research 20220701 7


<h4>Background</h4>Conventional flexible bronchoscopy has not achieved the high diagnostic yield for intrapulmonary lesions as seen with image-guided transthoracic biopsy. A thin convex probe endobronchial ultrasound bronchoscope (TCP-EBUS) with a 5.9-mm tip was designed to improve peripheral access over conventional EBUS bronchoscopes to facilitate real-time sampling of intrapulmonary lesions under ultrasound guidance.<h4>Methods</h4>TCP-EBUS was inserted into the distal airways of <i>ex-vivo</  ...[more]

Similar Datasets

| S-EPMC3981557 | biostudies-literature
| S-EPMC8575803 | biostudies-literature
| S-EPMC6961091 | biostudies-literature
| S-EPMC4693703 | biostudies-other
| S-EPMC5542957 | biostudies-other
| S-EPMC9840038 | biostudies-literature
| S-EPMC7927785 | biostudies-literature
| S-EPMC8264313 | biostudies-literature
| S-EPMC5380291 | biostudies-literature