Early Tibial Component Fractures in a Cementless, 3D-Printed, Titanium Implant.


ABSTRACT: Fracture of the tibial baseplate in total knee arthroplasty is a rare occurrence, particularly in short- and mid-term follow-up. This case series documents the first known report in the literature of fatigue fracture of a cementless, 3D-printed, highly porous titanium tibial component. We recommend regular follow-up visits with radiographs to confirm adequate total knee arthroplasty component positioning and alignment.

SUBMITTER: Lam AD 

PROVIDER: S-EPMC9576491 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Early Tibial Component Fractures in a Cementless, 3D-Printed, Titanium Implant.

Lam Alan D AD   Duffy Gavan P GP  

Arthroplasty today 20221012


Fracture of the tibial baseplate in total knee arthroplasty is a rare occurrence, particularly in short- and mid-term follow-up. This case series documents the first known report in the literature of fatigue fracture of a cementless, 3D-printed, highly porous titanium tibial component. We recommend regular follow-up visits with radiographs to confirm adequate total knee arthroplasty component positioning and alignment. ...[more]