ABSTRACT: Aim
This study was conducted to evaluate the accuracy of a newly developed multifrequency segmental (MFS) bioelectrical impedance analysis (BIA) method using an additional portable abdominal (PA) impedance analyzer, in the assessment of abdominal visceral fat area (VFA).Materials and methods
One hundred healthy Korean subjects aged 19 years or over (43 men and 57 women) were recruited, and VFA was estimated by a conventional MFS-BIA machine and a new MFS-BIA machine with a PA-BIA device, indicating MFS-VFA and MFS&PA-VFA, respectively. The accuracy of the VFA values was compared with those evaluated with CT at the level of the umbilicus (CT-VFA).Results
The mean age was 41 years and mean body mass index (BMI) was 24.4 kg/m2. The mean ± SD VFAs measured by CT, conventional MFS-BIA, and new MFS&PA-BIA together were 93.4 ± 60.9, 92.7 ± 53.4, and 93.6 ± 55.4 cm2, respectively. Correlation coefficients comparing CT-VFA with MFS-VFA and MFS&PA-VFA were 0.612 and 0.932, respectively (P < 0.001 for both). The mean difference between CT-VFA and MFS&PA-VFA was less affected by age, sex, and BMI compared with that between CT-VFA and MFS-VFA. Intraclass correlation coefficient (95% CI) between CT-VFA and MFS&PA-VFA was also greater than that between CT-VFA and MFS-VFA, 0.96 (0.95-0.98) vs. 0.76 (0.64-0.84), respectively.Conclusion
In this study, application of a newly developed MFS-BIA machine combined with a PA-BIA device significantly improved the correlation with CT-measured VFA without proportional error. This novel approach using advanced technology may be able to provide more reliable estimates of abdominal VFA.
SUBMITTER: Yoon JW
PROVIDER: S-EPMC9597369 | biostudies-literature | 2022
REPOSITORIES: biostudies-literature
Yoon Ji Won JW Sohn Minji M Moon Ji Hye JH Lim Soo S
Frontiers in nutrition 20221012
<h4>Aim</h4>This study was conducted to evaluate the accuracy of a newly developed multifrequency segmental (MFS) bioelectrical impedance analysis (BIA) method using an additional portable abdominal (PA) impedance analyzer, in the assessment of abdominal visceral fat area (VFA).<h4>Materials and methods</h4>One hundred healthy Korean subjects aged 19 years or over (43 men and 57 women) were recruited, and VFA was estimated by a conventional MFS-BIA machine and a new MFS-BIA machine with a PA-BIA ...[more]