Tenea Anda-Gabriela AG Dinu Cristina C Buica George-Octavian GO Vasile Gabriela-Geanina GG
Sensors (Basel, Switzerland) 20230117 3
The paper presents the validation of an electrochemical procedure for on-site Hg<sup>2+</sup> ions determination in wastewater samples using a modified carbon screen-printed electrode (SPE) with a complexing polymeric film based on poly(2,2'-(ethane-1,2-diylbis((2-(azulen-2-ylamino)-2-oxoethyl)azanediyl))diacetic acid) (poly<b>L</b>). Using metal ions accumulation in an open circuit followed by anodic stripping voltammetry, the SPE-poly<b>L</b> electrode presents a linear range in the range of 2 ...[more]