ABSTRACT: Objective
Pulse-wave velocity (PWV), a common measure of arterial stiffness, can be measured continuously and across multiple body sites using photoplethysmography (PPG). The objective was to determine whether a simple photoplethysmography PPG PWV method agrees with a referent device.Approach
Photoplethysmography heart-finger PWV (hfPWV) and heart-toe PWV (htPWV) were compared to oscillometric carotid-wrist PWV (cwPWV) and carotid-ankle PWV (caPWV) referent measurements, respectively. In 30 adults (24.6 ± 4.8 years, body mass index 25.2 ± 5.9 kg/m2, 18 female), three measurements were made: two supine baseline measurements (Base 1, Base 2) and one measurement (Tilt) 5 min after a modified head-up tilt test (mHUTT). Overall agreement and repeated measures agreement (change in PPG PWV from Base to Tilt vs. change in referent PWV from Base to Tilt) were calculated using linear mixed models. Agreement estimates were expressed as intra-class correlation coefficients (ICC).Main results
For hfPWV there was strong overall agreement (ICC: 0.77, 95%CI: 0.67-0.85), but negligible and non-significant repeated measures agreement (ICC: 0.10, 95%CI: -0.18 to 0.36). For htPWV, there was moderate overall agreement (ICC:0.50, 95%CI: 0.31-0.65) and strong repeated measures agreement (ICC: 0.81, 95%CI: 0.69-0.89).Significance
Photoplethysmography can continuously measure PWV at multiple arterial segments with moderate-strong overall agreement. While further work with upper-limb PPG PWV is needed, PPG can adequately capture acute changes in lower-limb PWV.
SUBMITTER: Zieff G
PROVIDER: S-EPMC9941627 | biostudies-literature | 2023
REPOSITORIES: biostudies-literature
Zieff Gabriel G Stone Keeron K Paterson Craig C Fryer Simon S Diana Jake J Blackwell Jade J Meyer Michelle L ML Stoner Lee L
Frontiers in cardiovascular medicine 20230207
<h4>Objective</h4>Pulse-wave velocity (PWV), a common measure of arterial stiffness, can be measured continuously and across multiple body sites using photoplethysmography (PPG). The objective was to determine whether a simple photoplethysmography PPG PWV method agrees with a referent device.<h4>Approach</h4>Photoplethysmography heart-finger PWV (hfPWV) and heart-toe PWV (htPWV) were compared to oscillometric carotid-wrist PWV (cwPWV) and carotid-ankle PWV (caPWV) referent measurements, respecti ...[more]