Cross-Calibration of iDXA and pQCT Scanners at Rural and Urban Research Sites in The Gambia, West Africa.


ABSTRACT: Between-scanner differences in measures of bone and body composition can obscure or exaggerate physiological differences in multi-site studies or the magnitude of changes in longitudinal studies. We conducted a cross-calibration study at two bone imaging centres in The Gambia, West Africa where DXA (dual-energy X-ray absorptiometry) and pQCT (peripheral Quantitative-Computed Tomography) are routinely used. Repeat scans were obtained from 64 Gambian adults (58% Male) aged Mean(SD) 30.9 (13.5) years with Mean(SD) body mass index (BMI) 21.7 (4.0) kg/m2, using DXA (GE Lunar iDXA, whole body [WB], total hip [TH], lumbar spine [LS]) and pQCT (Stratec XCT2000L/XCT2000, tibia 4%, 50% sites). Between-scanner differences were tested using paired t tests (p < 0.05). Between-scanner correlation was explored with linear regression, and cross-calibration equations derived. Bland-Altman analysis investigated machine trend/bias. When differences were detected (p < 0.05), cross-calibration equations were applied to urban values, with t tests and Bland Altman analysis repeated. Between-scanner differences exceeded the predefined level of statistical significance (p < 0.05) for WB aBMD and BA; all pQCT measures vBMD, BMC, cortical cross-sectional area (CSA) and stress-strain index (SSI). Between-scanner correlation was high (R2:0.92-0.99), except pQCT Mu.Den (R2 = 0.51). Bland Altman plots indicated bias increased with increasing BMD. Cross-calibration equations attenuated all between-scanner differences and systematic bias. Cross-calibration, particularly of pQCT scanners, is an important consideration in multi-site studies particularly where between population comparisons are intended. Our experiences and findings may be generalisable to other resource-limited settings where the logistics of sourcing parts and in-country repair may result in lengthy scanner downtime.

SUBMITTER: O Breasail M 

PROVIDER: S-EPMC9977475 | biostudies-literature | 2023 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Cross-Calibration of iDXA and pQCT Scanners at Rural and Urban Research Sites in The Gambia, West Africa.

Ó Breasail Mícheál M   Janha Ramatoulie R   Zengin Ayse A   Pearse Camille C   Jarjou Landing L   Prentice Ann A   Ward Kate A KA  

Calcified tissue international 20230302 5


Between-scanner differences in measures of bone and body composition can obscure or exaggerate physiological differences in multi-site studies or the magnitude of changes in longitudinal studies. We conducted a cross-calibration study at two bone imaging centres in The Gambia, West Africa where DXA (dual-energy X-ray absorptiometry) and pQCT (peripheral Quantitative-Computed Tomography) are routinely used. Repeat scans were obtained from 64 Gambian adults (58% Male) aged Mean(SD) 30.9 (13.5) yea  ...[more]

Similar Datasets

| S-EPMC6798247 | biostudies-literature
| S-EPMC2851933 | biostudies-literature
| S-EPMC2361674 | biostudies-literature
| S-EPMC5321282 | biostudies-literature
| S-EPMC7611936 | biostudies-literature
| S-EPMC3204775 | biostudies-literature
| S-EPMC4517717 | biostudies-literature
| S-EPMC11001577 | biostudies-literature
2015-10-26 | GSE53522 | GEO