Cryo-EM reconstruction of heavy chain mouse apoferritin, 8 datasets testing dose rate and correlative double sampling mode for the Gatan K3 direct electron detector. [1143 multi-frame micrographs composed of 120 frames in LZW compressed TIFF format


ABSTRACT:

SUBMITTER: Ming Sun 

PROVIDER: EMPIAR-10588 | biostudies-other |

REPOSITORIES: biostudies-other

altmetric image

Publications

Practical considerations for using K3 cameras in CDS mode for high-resolution and high-throughput single particle cryo-EM.

Sun Ming M   Azumaya Caleigh M CM   Tse Eric E   Bulkley David P DP   Harrington Matthew B MB   Gilbert Glenn G   Frost Adam A   Southworth Daniel D   Verba Kliment A KA   Cheng Yifan Y   Agard David A DA  

Journal of structural biology 20210511 3


Detector technology plays a pivotal role in high-resolution and high-throughput cryo-EM structure determination. Compared with the first-generation, single-electron counting direct detection camera (Gatan K2), the latest K3 camera is faster, larger, and now offers a correlated-double sampling mode (CDS). Importantly this results in a higher DQE and improved throughput compared to its predecessor. In this study, we focused on optimizing camera data collection parameters for daily use within a cry  ...[more]