Modes of failure of a threaded acetabular cup : a radiographic study with EBRA of 42 revised cups.


ABSTRACT: A radiographic study of a singular type of threaded cup, revised due to aseptic loosening, was performed with Einzel-Bild-Roentgen-Analyse (EBRA) to detect eventual risk factors and patterns of loosening. Fifty-three cups of 50 patients were revised. Forty-two cups could be measured with EBRA. No cup had obvious osteolysis, 33 cups demonstrated early migration, and all cups had migrated at the time of revision. Twenty-eight of the cups had a change of inclination and 21 of anteversion, respectively. The mean migration was 1.9 mm in the medial and 7.9 mm in the cranial direction; the mean wear rate was 0.2 mm/year. Cups with early migration had a higher migration rate. All male patients had early migration; medially placed cups had less migration than the other cups. Wear was not significantly affected by the migration of the implant. No cup had a complete radiolucent line, and the only radiographic sign for loosening was the change of position of the cup. Medial placement showed less migration in case of loosening. Regular radiographic follow-up is recommended for the examined implant, and the cup should not be used in the future.

SUBMITTER: Ilchmann T 

PROVIDER: S-EPMC2267560 | biostudies-other | 2007 Apr

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC2505130 | biostudies-other
| S-EPMC3851667 | biostudies-literature
| S-EPMC2269018 | biostudies-other
| S-EPMC6985010 | biostudies-literature
| S-EPMC5251254 | biostudies-literature
| S-EPMC8197737 | biostudies-literature
| S-EPMC5873012 | biostudies-literature
| S-EPMC6123169 | biostudies-literature
| S-EPMC7307242 | biostudies-literature
| S-EPMC6920715 | biostudies-literature