Sensitivity to interaural time differences with combined cochlear implant and acoustic stimulation.
Ontology highlight
ABSTRACT: The interaural time difference (ITD) is an important cue to localize sound sources. Sensitivity to ITD was measured in eight users of a cochlear implant (CI) in the one ear and a hearing aid (HA) in the other severely impaired ear. The stimulus consisted of an electric pulse train of 100 pps and an acoustic filtered click train. Just-noticeable differences (JNDs) in ITD were measured using a lateralization paradigm. Four subjects exhibited median JNDs in ITD of 156, 341, 254, and 91 mus; the other subjects could not lateralize the stimuli consistently. Only the subjects who could lateralize had average acoustic hearing thresholds at 1,000 and 2,000 Hz better than 100-dB SPL. The electric signal had to be delayed by 1.5 ms to achieve synchronous stimulation at the auditory nerves.