Unknown

Dataset Information

0

Subharmonic analysis using singular-value decomposition of ultrasound contrast agents.


ABSTRACT: Ultrasound contrast agents (UCAs) are designed to be used below 10 MHz, but interest is growing in studying the response of agents to high-frequency ultrasound. In this study, the subharmonic response of polymer-shelled UCAs with a mean diameter of 1.1 mum excited with 40-MHz tone-bursts of 1-20 cycles was analyzed. UCAs were diluted in water and streamed through a flow phantom that permitted single-bubble backscatter events to be acquired at peak-negative pressures from 0.75 to 5.0 MPa. At each exposure condition, 1000 single-bubble-backscatter events were digitized. Subharmonic content at 20 MHz was screened using a conventional and a singular-value-decomposition (SVD) method. The conventional method evaluated each event spectrum individually while the SVD method treated the 1000-event data set at one time. A subharmonic score (SHS) indicative of how much subharmonic content a 1000-event data set contained was computed for both methods. Empirical-simulation results indicated that SHSs obtained from the two methods were linearly related. Also, experimental data with both methods indicated that subharmonic likelihood increased with pulse duration and peaked near 2 MPa. The SVD method also yielded quantitative information about subharmonic events not available with the conventional method.

SUBMITTER: Mamou J 

PROVIDER: S-EPMC2719484 | biostudies-other | 2009 Jun

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC263735 | biostudies-literature
| S-EPMC2562393 | biostudies-literature
| S-EPMC6361234 | biostudies-literature
| S-EPMC10407287 | biostudies-literature
| S-EPMC7385012 | biostudies-literature
| S-EPMC5831701 | biostudies-literature
| S-EPMC7753369 | biostudies-literature
| S-EPMC6389416 | biostudies-literature
| S-EPMC8054608 | biostudies-literature
| S-EPMC2397332 | biostudies-literature