Unknown

Dataset Information

0

Design of malaria diagnostic criteria for the Sysmex XE-2100 hematology analyzer.


ABSTRACT: Thick film, the standard diagnostic procedure for malaria, is not always ordered promptly. A failsafe diagnostic strategy using an XE-2100 analyzer is proposed, and for this strategy, malaria diagnostic models for the XE-2100 were developed and tested for accuracy. Two hundred eighty-one samples were distributed into Plasmodium vivax, P. falciparum, and acute febrile syndrome groups for model construction. Model validation was performed using 60% of malaria cases and a composite control group of samples from AFS and healthy participants from endemic and non-endemic regions. For P. vivax, two observer-dependent models (accuracy = 95.3-96.9%), one non-observer-dependent model using built-in variables (accuracy = 94.7%), and one non-observer-dependent model using new and built-in variables (accuracy = 96.8%) were developed. For P. falciparum, two non-observer-dependent models (accuracies = 85% and 89%) were developed. These models could be used by health personnel or be integrated as a malaria alarm for the XE-2100 to prompt early malaria microscopic diagnosis.

SUBMITTER: Campuzano-Zuluaga G 

PROVIDER: S-EPMC2829900 | biostudies-other | 2010 Mar

REPOSITORIES: biostudies-other

altmetric image

Publications


Thick film, the standard diagnostic procedure for malaria, is not always ordered promptly. A failsafe diagnostic strategy using an XE-2100 analyzer is proposed, and for this strategy, malaria diagnostic models for the XE-2100 were developed and tested for accuracy. Two hundred eighty-one samples were distributed into Plasmodium vivax, P. falciparum, and acute febrile syndrome groups for model construction. Model validation was performed using 60% of malaria cases and a composite control group of  ...[more]