Unknown

Dataset Information

0

Comparison of phase-shifting techniques for in vivo full-range, high-speed Fourier-domain optical coherence tomography.


ABSTRACT: Single spectrometer-based complex conjugate artifact removal methods are evaluated for in vivo imaging with complementary metal-oxide semiconductor line scan camera based high-speed Fourier-domain optical coherence tomography (FD-OCT) at 100,000 axial scans per second. Performance of three different phase-shifting methods with the same OCT engine is evaluated using modified data acquisition schemes, depending on the requirements of each phase-shifting technique. The suppression ratio of complex conjugate artifact images using a paperboard is assessed for all tested methods. Several other characteristics, including a list of additional hardware requirements (beyond standard FD-OCT components) and data acquisition schemes for each of the methods is presented. In vivo full-range images of human fingerpad and nail are shown and compared with standard FD-OCT images. Additionally, a complex-conjugate-free human retinal volume acquired at the speed of 100,000 A-scans/s is presented.

SUBMITTER: Kim DY 

PROVIDER: S-EPMC2973988 | biostudies-other | 2010 Sep-Oct

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC8452642 | biostudies-literature
| S-EPMC5894866 | biostudies-literature
| S-EPMC3292386 | biostudies-literature
| S-EPMC3144967 | biostudies-other
| S-EPMC2136433 | biostudies-literature
| S-EPMC3005697 | biostudies-other