Convenient quantification of methanol concentration detection utilizing an integrated microfluidic chip.
Ontology highlight
ABSTRACT: A rapid and simple technique is proposed for methanol concentration detection using a PMMA (Polymethyl-Methacrylate) microfluidic chip patterned using a commercially available CO2 laser scriber. In the proposed device, methanol and methanol oxidase (MOX) are injected into a three-dimensional circular chamber and are mixed via a vortex stirring effect. The mixture is heated to prompt the formation of formaldehyde and is flowed into a rectangular chamber, to which fuchsin-sulphurous acid is then added. Finally, the microchip is transferred to a UV spectrophotometer for methanol detection purposes. The experimental results show that a correlation coefficient of R(2) = 0.9940 is obtained when plotting the optical density against the methanol concentration for samples and an accuracy as high as 93.1% are compared with the determined by the high quality gas chromatography with concentrations in the range of 2 ∼ 100 ppm. The methanol concentrations of four commercial red wines are successfully detected using the developed device. Overall, the results show that the proposed device provides a rapid and accurate means of detecting the methanol concentration for a variety of applications in the alcoholic beverage inspection and control field.
SUBMITTER: Wang YN
PROVIDER: S-EPMC3432083 | biostudies-other | 2012 Sep
REPOSITORIES: biostudies-other
ACCESS DATA