Treatment of intracranial atherosclerotic disease with a balloon-expandable paclitaxel eluting stent: procedural safety, efficacy and mid-term patency.


ABSTRACT: Procedural safety and high rates of in-stent recurrent stenotic lesions (ISR) remain a concern in the endovascular treatment of intracranial atherosclerotic disease (ICAD). In the present study technical feasibility, safety and efficacy of the paclitaxel eluting balloon-expandable coronary stent Coroflex(®) Please was assessed in the treatment of ICAD.A total of 95 patients (79 male; median age 68 years) with 106 intracranial atherosclerotic stenotic lesions underwent endovascular treatment using Coroflex(®) Please stents (B. Braun, Melsungen, Germany). Location and degree of target stenoses before and after treatment and at follow-up and adverse clinical sequelae of treatment were registered. Post-procedural medication included 100 mg acetylsalicylic acid (ASA) and 75 mg clopidogrel for 1 year. Angiographic follow-up was scheduled for 6, 12, 26 and 52 weeks after the treatment.The lesion locations were as follows: internal carotid artery (ICA) petrous (n = 44, 42%), ICA cavernous (n = 43, 41%), ICA paraclinoid (n = 4, 4%), intradural vertebral artery (VA; n = 11, 10%) and basilar artery (BA; n = 4, 4%). Of the lesions seven could not be treated due to difficult anatomy and stent stiffness (7% technical failure rate). The combined post-interventional neurological morbidity and mortality rate, including stroke, intracerebral hemorrhage (ICH), subarachnoid hemorrhage (SAH) and carotid cavernous fistula (CCF) was n = 4 (3.7%) within and n = 1 (0.9%) at and beyond 30 days, respectively. Angiographic and clinical follow-up examinations were carried out for 78 (78%) of the lesions (mean 16.1 months, maximum 48 months). Asymptomatic recurrent stenosis was seen in 3 out of 78 (3.8%) lesions and there was 1 case of late stent thrombosis (0.9%).Treatment of ICAD using drug-eluting coronary stents is safe and effective but technical failure due to stent stiffness remains a problem. Application of the more flexible, newest generation thin-strut stents, however, shows promising results.

SUBMITTER: Vajda Z 

PROVIDER: S-EPMC3432207 | biostudies-other | 2012 Sep

REPOSITORIES: biostudies-other

altmetric image

Publications

Treatment of intracranial atherosclerotic disease with a balloon-expandable paclitaxel eluting stent: procedural safety, efficacy and mid-term patency.

Vajda Z Z   Aguilar M M   Göhringer T T   Horváth-Rizea D D   Bäzner H H   Henkes H H  

Clinical neuroradiology 20120118 3


<h4>Purpose</h4>Procedural safety and high rates of in-stent recurrent stenotic lesions (ISR) remain a concern in the endovascular treatment of intracranial atherosclerotic disease (ICAD). In the present study technical feasibility, safety and efficacy of the paclitaxel eluting balloon-expandable coronary stent Coroflex(®) Please was assessed in the treatment of ICAD.<h4>Methods</h4>A total of 95 patients (79 male; median age 68 years) with 106 intracranial atherosclerotic stenotic lesions under  ...[more]

Similar Datasets

| S-EPMC5485564 | biostudies-literature
| S-EPMC7431892 | biostudies-literature
| S-EPMC4428173 | biostudies-literature
| S-EPMC4920989 | biostudies-literature
| S-EPMC5839641 | biostudies-literature
| S-EPMC5850679 | biostudies-literature
| S-EPMC7336236 | biostudies-literature
| S-EPMC5837867 | biostudies-literature