Unknown

Dataset Information

0

Improved Bloch-Siegert based B1 mapping by reducing off-resonance shift.


ABSTRACT: An MRI method based on the Bloch-Siegert (BS) shift phenomenon was recently proposed as a fast and precise way to map a radio frequency (RF) transmit field (B1(+) field). For MRI at high field, the mapping sensitivity of this approach was limited by tissue heating associated with a BS irradiation pulse. To mitigate this, we investigated the possibility of lowering the off-resonance frequency of this pulse since theoretical analysis indicated that the sensitivity of Bloch-Siegert based B1(+) mapping could be substantially improved when irradiating closer to resonance. Using optimized irradiation pulse shape and gradient crushers to minimize direct excitation effects, in vivo experiments on human brains at 7 T confirmed improved sensitivity with this approach. Improved sensitivity translated into an 80% reduction in B1(+) estimation errors without increasing tissue heating. Published 2013. This article is a U.S. Government work and is in the public domain in the USA.

SUBMITTER: Duan Q 

PROVIDER: S-EPMC3669656 | biostudies-other | 2013 Sep

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC6771691 | biostudies-literature
| S-EPMC5076794 | biostudies-literature
| S-EPMC4169365 | biostudies-literature
| S-EPMC6109041 | biostudies-literature
| S-EPMC5332527 | biostudies-literature
| S-EPMC4070276 | biostudies-other
| S-EPMC2916486 | biostudies-literature
| S-EPMC3085061 | biostudies-literature
| S-EPMC7083691 | biostudies-literature
| S-EPMC6354937 | biostudies-literature