Unknown

Dataset Information

0

Infrared mapping of ultrasound fields generated by medical transducers: feasibility of determining absolute intensity levels.


ABSTRACT: Considerable progress has been achieved in the use of infrared (IR) techniques for qualitative mapping of acoustic fields of high intensity focused ultrasound (HIFU) transducers. The authors have previously developed and demonstrated a method based on IR camera measurement of the temperature rise induced in an absorber less than 2 mm thick by ultrasonic bursts of less than 1 s duration. The goal of this paper was to make the method more quantitative and estimate the absolute intensity distributions by determining an overall calibration factor for the absorber and camera system. The implemented approach involved correlating the temperature rise measured in an absorber using an IR camera with the pressure distribution measured in water using a hydrophone. The measurements were conducted for two HIFU transducers and a flat physiotherapy transducer of 1 MHz frequency. Corresponding correction factors between the free field intensity and temperature were obtained and allowed the conversion of temperature images to intensity distributions. The system described here was able to map in good detail focused and unfocused ultrasound fields with sub-millimeter structure and with local time average intensity from below 0.1 W/cm(2) to at least 50 W/cm(2). Significantly higher intensities could be measured simply by reducing the duty cycle.

SUBMITTER: Khokhlova VA 

PROVIDER: S-EPMC3745508 | biostudies-other | 2013 Aug

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC7059213 | biostudies-literature
| S-EPMC6077835 | biostudies-other
| S-EPMC2600872 | biostudies-literature
| S-EPMC8433336 | biostudies-literature
| S-EPMC7032074 | biostudies-literature
| S-EPMC6376189 | biostudies-literature
| S-EPMC5471247 | biostudies-literature
| S-EPMC4221264 | biostudies-literature
| S-EPMC8693467 | biostudies-literature
| S-EPMC3610400 | biostudies-literature