Unknown

Dataset Information

0

Enhanced detection efficiency of direct conversion X-ray detector using polyimide as hole-blocking layer.


ABSTRACT: In this article we demonstrate the performance of a direct conversion amorphous selenium (a-Se) X-ray detector using biphenyldisnhydride/1,4 phenylenediamine (BPDA/PPD) polyimide (PI) as a hole-blocking layer. The use of a PI layer with a-Se allows detector operation at high electric fields (≥10 V/μm) while maintaining low dark current, without deterioration of transient performance. The hole mobility of the PI/a-Se device is measured by the time-of-flight method at different electric fields to investigate the effect of the PI layer on detector performance. It was found that hole mobility as high as 0.75 cm(2)/Vs is achievable by increasing the electric field in the PI/a-Se device structure. Avalanche multiplication is also shown to be achievable when using PI as a blocking layer. Increasing the electric field within a-Se reduces the X-ray ionization energy, increases hole mobility, and improves the dynamic range and sensitivity of the detector.

SUBMITTER: Abbaszadeh S 

PROVIDER: S-EPMC3842535 | biostudies-other | 2013

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC8948920 | biostudies-literature
| S-EPMC4010259 | biostudies-literature
| S-EPMC7407640 | biostudies-literature
| S-EPMC4315901 | biostudies-literature
| S-EPMC9403580 | biostudies-literature
| S-EPMC9146946 | biostudies-literature
| S-EPMC7730663 | biostudies-literature
| S-EPMC7794468 | biostudies-literature
| S-EPMC3543372 | biostudies-other
| S-EPMC7240709 | biostudies-literature