Kinetics of arterial carbon dioxide during veno-venous extracorporeal membrane oxygenation support in an apnoeic porcine model.


ABSTRACT: Extracorporeal membrane oxygenation (ECMO) is a technique widely used worldwide to improve gas exchange. Changes in ECMO settings affect both oxygen and carbon dioxide. The impact on oxygenation can be followed closely by continuous pulse oximeter. Conversely, carbon dioxide equilibrates much slower and is not usually monitored directly.We investigated the time to stabilization of arterial carbon dioxide partial pressure (PaCO2) following step changes in ECMO settings in 5 apnoeic porcine models under veno-venous ECMO support with polymethylpentene membranes. We collected sequential arterial blood gases at a pre-specified interval of 50 min using a sequence of standardized blood and sweep gas flow combinations.Following the changes in ECMO parameters, the kinetics of carbon dioxide was dependent on sweep gas and ECMO blood flow. With a blood flow of 1500 mL/min, PaCO2 takes longer than 50 min to equilibrate following the changes in sweep gas flow. Furthermore, the sweep gas flow from 3.0 to 10.0 L/min did not significantly affect PaCO2. However, with a blood flow of 3500 mL/min, 50 min was enough for PaCO2 to reach the equilibrium and every increment of sweep gas flow (up to 10.0 L/min) resulted in additional reductions of PaCO2.Fifty minutes was enough to reach the equilibrium of PaCO2 after ECMO initiation or after changes in blood and sweep gas flow with an ECMO blood flow of 3500 ml/min. Longer periods may be necessary with lower ECMO blood flows.

SUBMITTER: Mendes PV 

PROVIDER: S-EPMC4703593 | biostudies-other | 2016 Dec

REPOSITORIES: biostudies-other

altmetric image

Publications

Kinetics of arterial carbon dioxide during veno-venous extracorporeal membrane oxygenation support in an apnoeic porcine model.

Mendes Pedro Vitale PV   Park Marcelo M   Maciel Alexandre Toledo AT   E Silva Débora Prudêncio DP   Friedrich Natalia N   Barbosa Edzangela Vasconcelos Santos EV   Hirota Adriana Sayuri AS   Schettino Guilherme Pinto Paula GP   Azevedo Luciano Cesar Pontes LC   Costa Eduardo Leite Vieira EL  

Intensive care medicine experimental 20160106 1


<h4>Background</h4>Extracorporeal membrane oxygenation (ECMO) is a technique widely used worldwide to improve gas exchange. Changes in ECMO settings affect both oxygen and carbon dioxide. The impact on oxygenation can be followed closely by continuous pulse oximeter. Conversely, carbon dioxide equilibrates much slower and is not usually monitored directly.<h4>Methods</h4>We investigated the time to stabilization of arterial carbon dioxide partial pressure (PaCO2) following step changes in ECMO s  ...[more]

Similar Datasets

| S-EPMC11292066 | biostudies-literature
| S-EPMC8300043 | biostudies-literature
| S-EPMC7230450 | biostudies-literature
| S-EPMC7316147 | biostudies-literature
| S-EPMC10894441 | biostudies-literature
| S-EPMC5542924 | biostudies-other
| S-EPMC8249841 | biostudies-literature
| S-EPMC7393440 | biostudies-literature
| S-EPMC8119573 | biostudies-literature
| S-EPMC6649222 | biostudies-literature