Molecular Form Differences Between Prostate-Specific Antigen (PSA) Standards Create Quantitative Discordances in PSA ELISA Measurements.
Ontology highlight
ABSTRACT: The prostate-specific antigen (PSA) assays currently employed for the detection of prostate cancer (PCa) lack the specificity needed to differentiate PCa from benign prostatic hyperplasia and have high false positive rates. The PSA calibrants used to create calibration curves in these assays are typically purified from seminal plasma and contain many molecular forms (intact PSA and cleaved subforms). The purpose of this study was to determine if the composition of the PSA molecular forms found in these PSA standards contribute to the lack of PSA test reliability. To this end, seminal plasma purified PSA standards from different commercial sources were investigated by western blot (WB) and in multiple research grade PSA ELISAs. The WB results revealed that all of the PSA standards contained different mass concentrations of intact and cleaved molecular forms. Increased mass concentrations of intact PSA yielded higher immunoassay absorbance values, even between lots from the same manufacturer. Standardization of seminal plasma derived PSA calibrant molecular form mass concentrations and purification methods will assist in closing the gaps in PCa testing measurements that require the use of PSA values, such as the % free PSA and Prostate Health Index by increasing the accuracy of the calibration curves.
SUBMITTER: McJimpsey EL
PROVIDER: S-EPMC4766849 | biostudies-other | 2016 Feb
REPOSITORIES: biostudies-other
ACCESS DATA