Ti-Ag-Pd alloy with good mechanical properties and high potential for biological applications.
Ontology highlight
ABSTRACT: Ti-based alloys containing Ag were produced by tilt-casting method and their properties were studied. Even in its as-cast state, Ti94Ag3Pd3 showed relatively high tensile properties, good electrochemical behavior, and good biocompatibility. The relatively good mechanical properties of the as-cast α-Ti-type Ti94Ag3Pd3 alloy (tensile strength up to 850 MPa and elongation of ~10%) can be explained by its severely deformed, fine crystalline structure. The high biocompatibility of Ti94Ag3Pd3 can be explained by the Ag-Pd interaction, which inhibits the release of Ag ions from the surface. Ag, in combination with Pd has no toxic effects and demonstrates useful antimicrobial properties. The Ti94Ag3Pd3 alloy shows a good potential to be applied as a biomedical implant alloy.
SUBMITTER: Zadorozhnyy VY
PROVIDER: S-EPMC4848501 | biostudies-other | 2016 Apr
REPOSITORIES: biostudies-other
ACCESS DATA