Unknown

Dataset Information

0

High-contrast grating resonators for label-free detection of disease biomarkers.


ABSTRACT: A label-free optical biosensor is described that employs a silicon-based high-contrast grating (HCG) resonator with a spectral linewidth of ~500 pm that is sensitive to ligand-induced changes in surface properties. The device is used to generate thermodynamic and kinetic data on surface-attached antibodies with their respective antigens. The device can detect serum cardiac troponin I, a biomarker of cardiac disease to 100 pg/ml within 4 mins, which is faster, and as sensitive as current enzyme-linked immuno-assays for cTnI.

SUBMITTER: Sun T 

PROVIDER: S-EPMC4893738 | biostudies-other | 2016 Jun

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC3729447 | biostudies-literature
| S-EPMC2800857 | biostudies-literature
| S-EPMC3367093 | biostudies-literature
| S-EPMC4987677 | biostudies-literature
| S-EPMC4642504 | biostudies-literature
| S-EPMC6059845 | biostudies-other
| S-EPMC8363218 | biostudies-literature
| S-EPMC6299719 | biostudies-literature
| S-EPMC4466219 | biostudies-literature
| S-EPMC5437060 | biostudies-literature