Accuracy of the Fluorescence-Activated Cell Sorting Assay for the Aquaporin-4 Antibody (AQP4-Ab): Comparison with the Commercial AQP4-Ab Assay Kit.


ABSTRACT: BACKGROUND:The aquaporin-4 antibody (AQP4-Ab) is a disease-specific autoantibody to neuromyelitis optica (NMO). We aimed to evaluate the accuracy of the FACS assay in detecting the AQP4-Ab compared with the commercial cell-based assay (C-CBA) kit. METHODS:Human embryonic kidney-293 cells were transfected with human aquaporin-4 (M23) cDNA. The optimal cut off values of FACS assay was tested using 1123 serum samples from patients with clinically definite NMO, those at high risk for NMO, patients with multiple sclerosis, patients with other idiopathic inflammatory demyelinating diseases, and negative controls. The accuracy of FACS assay and C-CBA were compared in consecutive 225 samples that were collected between January 2014 and June 2014. RESULTS:With a cut-off value of MFIi of 3.5 and MFIr of 2.0, the receiver operating characteristic curve for the FACS assay showed an area under the curve of 0.876. Among 225 consecutive sera, the FACS assay and C-CBA had a sensitivity of 77.3% and 69.7%, respectively, in differentiating the sera of definite NMO patients from sera of controls without IDD or of MS. Both assay had a good specificity of 100% in it. The overall positivity of the C-CBA among FACS-positive sera was 81.5%; moreover, its positivity was low as 50% among FACS-positive sera with relatively low MFIis. CONCLUSIONS:Both the FACS assay and C-CBA are sensitive and highly specific assays in detecting AQP4-Ab. However, in some sera with relatively low antibody titer, FACS-assay can be a more sensitive assay option. In real practice, complementary use of FACS assay and C-CBA will benefit the diagnosis of NMO patients, because the former can be more sensitive among low titer sera and the latter are easier to use therefore can be widely used.

SUBMITTER: Yang J 

PROVIDER: S-EPMC5033450 | biostudies-other | 2016

REPOSITORIES: biostudies-other

altmetric image

Publications

Accuracy of the Fluorescence-Activated Cell Sorting Assay for the Aquaporin-4 Antibody (AQP4-Ab): Comparison with the Commercial AQP4-Ab Assay Kit.

Yang Jiwon J   Kim Sung Min SM   Kim Yoo-Jin YJ   Cheon So Young SY   Kim Boram B   Jung Kyeong Cheon KC   Park Kyung Seok KS  

PloS one 20160922 9


<h4>Background</h4>The aquaporin-4 antibody (AQP4-Ab) is a disease-specific autoantibody to neuromyelitis optica (NMO). We aimed to evaluate the accuracy of the FACS assay in detecting the AQP4-Ab compared with the commercial cell-based assay (C-CBA) kit.<h4>Methods</h4>Human embryonic kidney-293 cells were transfected with human aquaporin-4 (M23) cDNA. The optimal cut off values of FACS assay was tested using 1123 serum samples from patients with clinically definite NMO, those at high risk for  ...[more]

Similar Datasets

| S-EPMC2805761 | biostudies-literature
| S-EPMC10116420 | biostudies-literature
| S-EPMC6787171 | biostudies-literature
2022-01-20 | GSE167944 | GEO
| S-EPMC3890327 | biostudies-literature
| S-EPMC4193741 | biostudies-literature
| S-EPMC2832968 | biostudies-literature
| S-EPMC4127988 | biostudies-literature
| S-EPMC3829826 | biostudies-literature
| S-EPMC4202686 | biostudies-literature