Unknown

Dataset Information

0

Multi-Echo-Based Echo-Planar Spectroscopic Imaging Using a 3T MRI Scanner.


ABSTRACT: The use of spin-echoes has been employed in an Echo-Planar Spectroscopic Imaging (EPSI) sequence to collect multiple phase encoded lines within a single TR in a Multi-Echo-based Echo-Planar Spectroscopic Imaging technique (MEEPSI). Despite the T₂ dependence on the amplitude of the spin-echoes, the Full Width at Half Maximum (FWHM) of the derived multi-echo Point Spread Function (PSF) is shown to decrease, indicating an improved overall spatial resolution without requiring any additional scan time. The improved spatial resolution is demonstrated in the one-dimensional (1D) spatial profiles of the N-Acetyl Aspartate (NAA) singlet along the phase encode dimension in a gray matter phantom. Although the improved spatial resolution comes at the expense of spectral resolution, it is shown in vivo that peak broadening due to T₂* decay is more significant than the loss of resolution from using spin-echoes and therefore does not affect the ability to quantify metabolites using the LCModel fitting algorithm.

SUBMITTER: Furuyama JK 

PROVIDER: S-EPMC5448876 | biostudies-other | 2011 Oct

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC3429710 | biostudies-literature
| S-EPMC4160415 | biostudies-other
| S-EPMC5725420 | biostudies-literature
| S-EPMC5836719 | biostudies-literature
| S-EPMC3662903 | biostudies-other
| S-EPMC5803468 | biostudies-literature
| S-EPMC9300143 | biostudies-literature
| S-EPMC4162886 | biostudies-literature
| S-EPMC3957332 | biostudies-literature
| S-EPMC4414683 | biostudies-literature