Unknown

Dataset Information

0

Measurement of the blood flow rate and velocity in coronary artery stenosis using intracoronary frequency domain optical coherence tomography: Validation against fractional flow reserve.


ABSTRACT: The main objective of this study was to assess the blood flow rate and velocity in coronary artery stenosis using intracoronary frequency domain optical coherence tomography (FD-OCT). A correlation between fractional flow reserve (FFR) and FD-OCT derived blood flow velocity is also included in this study.A total of 20 coronary stenoses in 15 patients were assessed consecutively by quantitative coronary angiography (QCA), FFR and FD-OCT. A percutaneous coronary intervention (PCI) optimization system was used in this study which combines wireless FFR measurement and FD-OCT imaging in one platform. Stenoses were labelled severe if FFR ≤ 0.8. Blood flow rate and velocity in each stenosis segment were derived from the volumetric analysis of the FD-OCT pull back images. The FFR value was ≤ 0.80 in 5 stenoses (25%). The mean blood flow rate in severe coronary stenosis (n = 5) was 2.54 ± 0.55 ml/s as compared to 4.81 ± 1.95 ml/s in stenosis with FFR > 0.8 (n = 15). A good and significant correlation between FFR and FD-OCT blood flow velocity in coronary artery stenosis (r = 0.74, p < 0.001) was found.The assessment of stenosis severity using FD-OCT derived blood flow rate and velocity has the ability to overcome many limitations of QCA and intravascular ultrasound (IVUS).

SUBMITTER: Zafar H 

PROVIDER: S-EPMC5497166 | biostudies-other | 2014 Dec

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC6601174 | biostudies-literature
| S-EPMC6159406 | biostudies-other
| S-EPMC7686372 | biostudies-literature
| S-EPMC5850206 | biostudies-literature
| S-EPMC6070545 | biostudies-literature
| S-EPMC7525716 | biostudies-literature
| S-EPMC9234158 | biostudies-literature
| S-EPMC8080642 | biostudies-literature
| S-EPMC5586447 | biostudies-literature
| S-EPMC6711086 | biostudies-other