Unknown

Dataset Information

0

Voxel Placement Precision for GABA-Edited Magnetic Resonance Spectroscopy.


ABSTRACT: The purpose of the present study was to assess the reproducibility of voxel placement for GABA-edited MRS. GABA-edited MRS data were acquired in 13 healthy volunteers from (3 cm)3 voxel; and within the same session a second acquisition was independently prescribed. A three-dimensional voxel mask image was reconstructed in T1-image-space using the SVMask tool (in house software). Reproducibility of voxel placement was assessed using the Dice overlap coefficient, both within-subject and between-subject following co-registration of T1 images and transformation of voxel mask images to standard space. Within-subject overlap coefficients were 86% ± 5%. Between-subject overlap coefficients were 75% ± 10%. For the two voxel locations considered (occipital and sensorimotor), voxel overlap was very similar. Between-subject values are higher due to between-session effects, anatomical variability and volume mismatch in standard space. While surprisingly low in terms of volume overlap, the overlap coefficients correspond to acceptable linear displacements.

SUBMITTER: Bai X 

PROVIDER: S-EPMC5497851 | biostudies-other | 2017 Mar

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC6007442 | biostudies-literature
| S-EPMC6291009 | biostudies-literature
| S-EPMC9203918 | biostudies-literature
2023-11-03 | GSE239379 | GEO
| S-EPMC7522573 | biostudies-literature
| S-EPMC5700835 | biostudies-literature
| S-EPMC7347431 | biostudies-literature
2023-04-25 | GSE225302 | GEO
| S-EPMC8404144 | biostudies-literature
| S-EPMC6818968 | biostudies-literature