Unknown

Dataset Information

0

METHODS FOR ESTIMATION OF BLOOD GLUCOSE : A COMPARATIVE EVALUATION.


ABSTRACT: Glucose in blood is the most frequent analyzed parameter in a clinical chemistry laboratory. In Armed Forces Laboratories, copper reduction method (Modified Folin Wu) is commonly used. Here we have compared this method as well as O-Toluidine and GOD-POD method with reference UV-Hexokinase method. Both Modified Folin Wu and O-Toluidine showed upward deviation with substantial imprecision (CV=4.9% - 3.5% and 6% - 5.8% respectively) and inaccuracy (Average deviation = 5.76% and 10.68% respectively). GOD-POD method was found linear (up to 500 mg/dl), with good precision (CV=0.7% to 1.4%) and accuracy (Average deviation= -0.97%). This method is simple, rapid, economical and sensitive, and can be adopted to a routine colorimeter.

SUBMITTER: Ambade VN 

PROVIDER: S-EPMC5531325 | biostudies-other | 1998 Apr

REPOSITORIES: biostudies-other

Similar Datasets

2020-11-30 | GSE150587 | GEO
| S-EPMC2238724 | biostudies-literature
| S-EPMC4983420 | biostudies-literature
2019-09-17 | GSE124167 | GEO
| S-EPMC6387385 | biostudies-literature
| S-EPMC5818632 | biostudies-literature
| S-EPMC9883484 | biostudies-literature
| S-EPMC5388484 | biostudies-literature
| S-EPMC4809601 | biostudies-literature
2020-10-01 | MSV000086224 | MassIVE