Zhao Tianyu T Cai Bin B Sun Baozhou B Grantham Kevin K Mutic Sasa S Klein Eric E
Journal of applied clinical medical physics 20150908 5
The purpose of this study was to evaluate the use of diverging-cut aperture to minimize collimator contamination in proton therapy. Two sets of apertures with nondivergent and divergent edge were fabricated to produce a 10 cm × 10 cm field at the radiation isocenter of a single-room proton therapy unit. Transverse profiles were acquired in a scanning water tank with both aperture sets. Up to 9.5% extra dose was observed from aperture scattering near the field edges with the nondivergent aperture ...[more]