Detection of artificial pulmonary lung nodules in ultralow-dose CT using an ex vivo lung phantom.


ABSTRACT: To assess the image quality of 3 different ultralow-dose CT protocols on pulmonary nodule depiction in a ventilated ex vivo-system.Four porcine lungs were inflated inside a dedicated chest phantom and prepared with n = 195 artificial nodules (0.5-1 mL). The artificial chest wall was filled with water to simulate the absorption of a human chest. Images were acquired with a 2x192-row detector CT using low-dose (reference protocol with a tube voltage of 120 kV) and 3 different ULD protocols (respective effective doses: 1mSv and 0.1mSv). A different tube voltage was used for each ULD protocol: 70kV, 100kV with tin filter (100kV_Sn) and 150kV with tin filter (150kV_Sn). Nodule delineation was assessed by two observers (scores 1-5, 1 = unsure, 5 = high confidence).The diameter of the 195 detected artificial nodules ranged from 0.9-21.5 mm (mean 7.84 mm ± 5.31). The best ULD scores were achieved using 100kV_Sn and 70 kV ULD protocols (4.14 and 4.06 respectively). Both protocols were not significantly different (p = 0.244). The mean score of 3.78 in ULD 150kV_Sn was significantly lower compared to the 100kV_Sn ULD protocol (p = 0.008).The results of this experiment, conducted in a realistic setting show the feasibility of ultralow-dose CT for the detection of pulmonary nodules.

SUBMITTER: Burgard CA 

PROVIDER: S-EPMC5752031 | biostudies-other | 2018

REPOSITORIES: biostudies-other

altmetric image

Publications

Detection of artificial pulmonary lung nodules in ultralow-dose CT using an ex vivo lung phantom.

Burgard Caroline Alexandra CA   Gaass Thomas T   Bonert Madeleine M   Bondesson David D   Thaens Natalie N   Reiser Maximilian Ferdinand MF   Dinkel Julien J  

PloS one 20180103 1


<h4>Objectives</h4>To assess the image quality of 3 different ultralow-dose CT protocols on pulmonary nodule depiction in a ventilated ex vivo-system.<h4>Materials and methods</h4>Four porcine lungs were inflated inside a dedicated chest phantom and prepared with n = 195 artificial nodules (0.5-1 mL). The artificial chest wall was filled with water to simulate the absorption of a human chest. Images were acquired with a 2x192-row detector CT using low-dose (reference protocol with a tube voltage  ...[more]

Similar Datasets

| S-EPMC5903960 | biostudies-literature
| S-EPMC7931546 | biostudies-literature
| S-EPMC7592972 | biostudies-literature
| S-EPMC6701577 | biostudies-literature
| S-EPMC7706885 | biostudies-literature
| S-EPMC5300187 | biostudies-literature
| S-EPMC7428081 | biostudies-literature
| S-EPMC3873253 | biostudies-literature
| S-EPMC7757901 | biostudies-literature
| S-EPMC4824919 | biostudies-literature