Geometric distortion of panoramic reconstruction in third molar tilting assessments: a comprehensive evaluation.


ABSTRACT: OBJECTIVES::To evaluate the geometric distortion of tilting of mandibular third molars with respect to second molars on panoramic reconstruction. METHODS::Cone-beam CT (CBCT) reconstructions of 160 third molars, obtained due to an indication of risk of inferior alveolar nerve damage during surgery, were used. CBCT-reconstructed panoramic images were used as bi-dimensional (2D) images, to avoid distortions other than geometric distortions. The angle between the second and the third molar was measured in 2D and three-dimensional (3D) images. Student's t-test was used to assess the null-hypothesis of no difference between 2D and 3D measurements. RESULTS::A significant mean difference (-2.3° ± 6.3°) between 2D and 3D measurements was found, with an absolute error of 3.6° ± 5.7° and a relative error of 10%. These findings comprehensively explain the geometric distortion on panoramic radiographs. CONCLUSIONS::Although a widely used and undoubtedly useful tool for diagnosis and surgical planning of mandibular third molar extractions, panoramic reconstruction are biased from geometric distortion that may influence surgical planning.

SUBMITTER: Lupi SM 

PROVIDER: S-EPMC6196059 | biostudies-other | 2018 Oct

REPOSITORIES: biostudies-other

altmetric image

Publications

Geometric distortion of panoramic reconstruction in third molar tilting assessments: a comprehensive evaluation.

Lupi Saturnino Marco SM   Galinetto Pietro P   Cislaghi Matteo M   Rodriguez Y Baena Arianna A   Scribante Andrea A   Rodriguez Y Baena Ruggero R  

Dento maxillo facial radiology 20180627 7


<h4>Objectives:</h4>To evaluate the geometric distortion of tilting of mandibular third molars with respect to second molars on panoramic reconstruction.<h4>Methods:</h4>Cone-beam CT (CBCT) reconstructions of 160 third molars, obtained due to an indication of risk of inferior alveolar nerve damage during surgery, were used. CBCT-reconstructed panoramic images were used as bi-dimensional (2D) images, to avoid distortions other than geometric distortions. The angle between the second and the third  ...[more]

Similar Datasets

| S-EPMC8844031 | biostudies-literature
| S-EPMC9547920 | biostudies-literature
| S-EPMC5750684 | biostudies-other
| S-EPMC8925877 | biostudies-literature
| S-EPMC9910288 | biostudies-literature
| S-EPMC10607494 | biostudies-literature
| 2184257 | ecrin-mdr-crc
| S-EPMC7304185 | biostudies-literature
| S-EPMC9376417 | biostudies-literature
| S-EPMC10465997 | biostudies-literature