Subjective and objective comparisons of image quality between ultra-high-resolution CT and conventional area detector CT in phantoms and cadaveric human lungs.


ABSTRACT: OBJECTIVES:To compare the image quality of the lungs between ultra-high-resolution CT (U-HRCT) and conventional area detector CT (AD-CT) images. METHODS:Image data of slit phantoms (0.35, 0.30, and 0.15 mm) and 11 cadaveric human lungs were acquired by both U-HRCT and AD-CT devices. U-HRCT images were obtained with three acquisition modes: normal mode (U-HRCTN: 896 channels, 0.5 mm × 80 rows; 512 matrix), super-high-resolution mode (U-HRCTSHR: 1792 channels, 0.25 mm × 160 rows; 1024 matrix), and volume mode (U-HRCTSHR-VOL: non-helical acquisition with U-HRCTSHR). AD-CT images were obtained with the same conditions as U-HRCTN. Three independent observers scored normal anatomical structures (vessels and bronchi), abnormal CT findings (faint nodules, solid nodules, ground-glass opacity, consolidation, emphysema, interlobular septal thickening, intralobular reticular opacities, bronchovascular bundle thickening, bronchiectasis, and honeycombing), noise, artifacts, and overall image quality on a 3-point scale (1 = worst, 2 = equal, 3 = best) compared with U-HRCTN. Noise values were calculated quantitatively. RESULTS:U-HRCT could depict a 0.15-mm slit. Both U-HRCTSHR and U-HRCTSHR-VOL significantly improved visualization of normal anatomical structures and abnormal CT findings, except for intralobular reticular opacities and reduced artifacts, compared with AD-CT (p < 0.014). Visually, U-HRCTSHR-VOL has less noise than U-HRCTSHR and AD-CT (p < 0.00001). Quantitative noise values were significantly higher in the following order: U-HRCTSHR (mean, 30.41), U-HRCTSHR-VOL (26.84), AD-CT (16.03), and U-HRCTN (15.14) (p < 0.0001). U-HRCTSHR and U-HRCTSHR-VOL resulted in significantly higher overall image quality than AD-CT and were almost equal to U-HRCTN (p < 0.0001). CONCLUSIONS:Both U-HRCTSHR and U-HRCTSHR-VOL can provide higher image quality than AD-CT, while U-HRCTSHR-VOL was less noisy than U-HRCTSHR. KEY POINTS:• Ultra-high-resolution CT (U-HRCT) can improve spatial resolution. • U-HRCT can reduce streak and dark band artifacts. • U-HRCT can provide higher image quality than conventional area detector CT. • In U-HRCT, the volume mode is less noisy than the super-high-resolution mode. • U-HRCT may provide more detailed information about the lung anatomy and pathology.

SUBMITTER: Yanagawa M 

PROVIDER: S-EPMC6223853 | biostudies-other | 2018 Dec

REPOSITORIES: biostudies-other

altmetric image

Publications

Subjective and objective comparisons of image quality between ultra-high-resolution CT and conventional area detector CT in phantoms and cadaveric human lungs.

Yanagawa Masahiro M   Hata Akinori A   Honda Osamu O   Kikuchi Noriko N   Miyata Tomo T   Uranishi Ayumi A   Tsukagoshi Shinsuke S   Tomiyama Noriyuki N  

European radiology 20180529 12


<h4>Objectives</h4>To compare the image quality of the lungs between ultra-high-resolution CT (U-HRCT) and conventional area detector CT (AD-CT) images.<h4>Methods</h4>Image data of slit phantoms (0.35, 0.30, and 0.15 mm) and 11 cadaveric human lungs were acquired by both U-HRCT and AD-CT devices. U-HRCT images were obtained with three acquisition modes: normal mode (U-HRCT<sub>N</sub>: 896 channels, 0.5 mm × 80 rows; 512 matrix), super-high-resolution mode (U-HRCT<sub>SHR</sub>: 1792 channels,  ...[more]

Similar Datasets

| S-EPMC2663638 | biostudies-other
| S-EPMC4711718 | biostudies-literature
| S-EPMC7248011 | biostudies-literature
| S-EPMC5570320 | biostudies-literature
| S-EPMC6849778 | biostudies-literature
| S-EPMC7365134 | biostudies-literature
| S-EPMC5550587 | biostudies-other
| S-EPMC6895079 | biostudies-literature
| S-EPMC6749483 | biostudies-literature
| S-EPMC8235100 | biostudies-literature