Ontology highlight
ABSTRACT:
SUBMITTER: Georgi C
PROVIDER: S-EPMC6393888 | biostudies-other | 2019
REPOSITORIES: biostudies-other
Georgi Christian C Neuß Michael M Möller Viviane V Seifert Martin M Butter Christian C
Case reports in cardiology 20190214
With a growing number of ICD recipients, device complications are seen more frequently in the clinical setting and outpatient departments. Among the most severe are ICD infections and inappropriate therapies caused by oversensing of atrial tachycardias or lead fracture. We report on a 76-year-old female patient with dilative cardiomyopathy and Broca's aphasia after stroke, who experienced 105 consecutive inappropriate ICD shocks due to cluster missensing of her fractured ICD lead. The diagnosis ...[more]