Low-Dose Imaging in a New Preclinical Total-Body PET/CT Scanner.


ABSTRACT: Ionizing radiation constitutes a health risk to imaging scientists and study animals. Both PET and CT produce ionizing radiation. CT doses in pre-clinical in vivo imaging typically range from 50 to 1,000 mGy and biological effects in mice at this dose range have been previously described. [18F]FDG body doses in mice have been estimated to be in the range of 100 mGy for [18F]FDG. Yearly, the average whole body doses due to handling of activity by PET technologists are reported to be 3-8 mSv. A preclinical PET/CT system is presented with design features which make it suitable for small animal low-dose imaging. The CT subsystem uses a X-source power that is optimized for small animal imaging. The system design incorporates a spatial beam shaper coupled with a highly sensitive flat-panel detector and very fast acquisition (<10 s) which allows for whole body scans with doses as low as 3 mGy. The mouse total-body PET subsystem uses a detector architecture based on continuous crystals, coupled to SiPM arrays and a readout based in rows and columns. The PET field of view is 150 mm axial and 80 mm transaxial. The high solid-angle coverage of the sample and the use of continuous crystals achieve a sensitivity of 9% (NEMA) that can be leveraged for use of low tracer doses and/or performing rapid scans. The low-dose imaging capabilities of the total-body PET subsystem were tested with NEMA phantoms, in tumor models, a mouse bone metabolism scan and a rat heart dynamic scan. The CT imaging capabilities were tested in mice and in a low contrast phantom. The PET low-dose phantom and animal experiments provide evidence that image quality suitable for preclinical PET studies is achieved. Furthermore, CT image contrast using low dose scan settings was suitable as a reference for PET scans. Total-body mouse PET/CT studies could be completed with total doses of <10 mGy.

SUBMITTER: Molinos C 

PROVIDER: S-EPMC6509903 | biostudies-other | 2019

REPOSITORIES: biostudies-other

altmetric image

Publications


Ionizing radiation constitutes a health risk to imaging scientists and study animals. Both PET and CT produce ionizing radiation. CT doses in pre-clinical <i>in vivo</i> imaging typically range from 50 to 1,000 mGy and biological effects in mice at this dose range have been previously described. [<sup>18</sup>F]FDG body doses in mice have been estimated to be in the range of 100 mGy for [<sup>18</sup>F]FDG. Yearly, the average whole body doses due to handling of activity by PET technologists are  ...[more]

Similar Datasets

| S-EPMC8241799 | biostudies-literature
| S-EPMC8455202 | biostudies-literature
| S-EPMC3751692 | biostudies-literature
| S-EPMC3350540 | biostudies-other
| S-EPMC8315067 | biostudies-literature
2010-06-09 | E-GEOD-1366 | biostudies-arrayexpress
| S-EPMC7807637 | biostudies-literature
| S-EPMC6072675 | biostudies-literature
2004-06-17 | GSE1366 | GEO
| S-EPMC6373185 | biostudies-literature