TNK1 BioID Proteomics Andersen
Ontology highlight
ABSTRACT: Qualitative LC MS/MS was performed on 4 uL of each sample, using a nanoAcquity UPLC system (Waters Corp) coupled to a Thermo Orbitrap Fusion Lumos high resolution accurate mass tandem mass spectrometer (Thermo) via a nanoelectrospray ionization source. Briefly, the sample was first trapped on a Symmetry C18 20 mm x 180 um trapping column (5 ul/min at 99.9/0.1 v/v water/acetonitrile), after which the analytical separation was performed using a 1.8 um Acquity HSS T3 C18 75 um x 250 mm column (Waters Corp.) with a 90-min linear gradient of 3 to 30% acetonitrile with 0.1% formic acid at a flow rate of 400 nanoliters/minute (nL/min) with a column temperature of 55C. Data collection on the Fusion Lumos mass spectrometer was performed in a data dependent acquisition (DDA) mode of acquisition with a r=120,000 (@ m/z 200) full MS scan from m/z 375 to 1500 with a target AGC value of 2e5 ions. MS/MS scans were acquired at Rapid scan rate (Ion Trap) with an AGC target of 5e3 ions and a max injection time of 200 ms. The total cycle time for MS and MS/MS scans was 2 sec. A 20 sec dynamic exclusion was employed to increase depth of coverage. The total analysis cycle time for each sample injection was approximately 2 hours.
INSTRUMENT(S): Orbitrap Fusion Lumos
ORGANISM(S): Homo Sapiens (ncbitaxon:9606)
SUBMITTER: Josh Anderson
PROVIDER: MSV000087618 | MassIVE | Mon Jun 14 10:05:00 BST 2021
REPOSITORIES: MassIVE