Impact of the used solvent on the reconstitution efficiency of evaporated biosamples for untargeted metabolomics studies


ABSTRACT:

Introduction

Untargeted metabolomics intends to objectively analyze a wide variety of compounds. Their diverse physicochemical properties make it difficult to choose an appropriate reconstitution solvent after sample evaporation without influencing the chromatography or hamper column sorbent integrity.

Objectives

The study aimed to identify the most appropriate reconstitution solvent for blood plasma samples in terms of feature recovery, four endogenous compounds, and one selected internal standard.

Methods

We investigated several reconstitution solvent mixtures containing acetonitrile and methanol to resolve human plasma extract and evaluated them concerning the peak areas of tryptophan-d5, glucose, creatinine, palmitic acid, and the phophatidylcholine PC(P-16:0/P-16:0), as well as the total feature count RESULTS: Results indicated that acetonitrile containing 30% methanol was best suited to match all tested criteria at least for human blood plasma samples.

Conclusion

Despite identifying the mixture of acetonitrile and methanol being suitable as solvent for human blood plasma extracts, we recommend to systematically test for an appropriate reconstitution solvent for each analyzed biomatrix.

INSTRUMENT(S): Liquid Chromatography MS - Negative (LC-MS (Negative)), Liquid Chromatography MS - Positive (LC-MS (Positive))

SUBMITTER: Sascha K. Manier 

PROVIDER: MTBLS1219 | MetaboLights | 2021-10-27

REPOSITORIES: MetaboLights

Dataset's files

Source:
altmetric image

Publications

Impact of the used solvent on the reconstitution efficiency of evaporated biosamples for untargeted metabolomics studies.

Manier Sascha K SK   Meyer Markus R MR  

Metabolomics : Official journal of the Metabolomic Society 20200302 3


<h4>Introduction</h4>Untargeted metabolomics intends to objectively analyze a wide variety of compounds. Their diverse physicochemical properties make it difficult to choose an appropriate reconstitution solvent after sample evaporation without influencing the chromatography or hamper column sorbent integrity.<h4>Objectives</h4>The study aimed to identify the most appropriate reconstitution solvent for blood plasma samples in terms of feature recovery, four endogenous compounds, and one selected  ...[more]

Similar Datasets

2023-01-17 | PXD038750 | Pride
2023-10-31 | GSE227777 | GEO
2017-07-23 | GSE59682 | GEO
2014-06-10 | GSE58303 | GEO
2014-06-10 | E-GEOD-58303 | biostudies-arrayexpress
2020-08-06 | MTBLS1465 | MetaboLights
2022-06-01 | GSE133348 | GEO
2022-04-14 | ST002152 | MetabolomicsWorkbench
2020-09-24 | GSE154567 | GEO
2022-02-22 | PXD023175 | Pride