Metabolomics

Dataset Information

0

Comparison of Full-Scan, Data-Dependent, and Data-Independent Acquisition Modes in Liquid Chromatography-Mass Spectrometry Based Untargeted Metabolomics


ABSTRACT: Full-scan, data-dependent acquisition (DDA), and data-independent acquisition (DIA) are the three common data acquisition modes in high resolution mass spectrometry-based untargeted metabolomics. It is an important yet underrated research topic on which acquisition mode is more suitable for a given untargeted metabolomics application. In this work, we compared the three data acquisition techniques using a standard mixture of 134 endogenous metabolites and a human urine sample. Both hydrophilic interaction and reversed-phase liquid chromatographic separation along with positive and negative ionization modes were tested. Both the standard mixture and urine samples generated consistent results. Full-scan mode is able to capture the largest number of metabolic features, followed by DIA and DDA (53.7% and 64.8% respective features fewer on average in urine than full-scan). Comparing the MS2 spectra in DIA and DDA, spectra quality is higher in DDA with average dot product score 83.1% higher than DIA in Urine(H), and the number of MS2 spectra (spectra quantity) is larger in DIA (on average 97.8% more than DDA in urine). Moreover, a comparison of relative standard deviation distribution between modes shows consistency in the quantitative precision, with the exception of DDA showing a minor disadvantage (on average 19.8% and 26.8% fewer features in urine with RSD < 5% than full-scan and DIA). In terms of data preprocessing convenience, full-scan and DDA data can be processed by well-established software. In contrast, several bioinformatic issues remain to be addressed in processing DIA data and the development of more effective computational programs is highly demanded.

INSTRUMENT(S): Liquid Chromatography MS - Reverse Phase (POS)” Liquid Chromatography MS - HILIC (NEG)”, Liquid Chromatography MS - HILIC (NEG), Liquid Chromatography MS - Reverse Phase (POS), Liquid Chromatography MS - Positive (LC-MS (Positive))

SUBMITTER: Jian Guo 

PROVIDER: MTBLS1572 | MetaboLights | 2020-05-21

REPOSITORIES: MetaboLights

Dataset's files

Source:
Action DRS
MTBLS1572 Other
FILES Other
a_MTBLS1572_StdMixHigh_NEG_LC-MS.txt Txt
a_MTBLS1572_StdMixHigh_POS_LC-MS.txt Txt
a_MTBLS1572_StdMixLow_NEG_LC-MS.txt Txt
Items per page:
1 - 5 of 16
altmetric image

Publications

Comparison of Full-Scan, Data-Dependent, and Data-Independent Acquisition Modes in Liquid Chromatography-Mass Spectrometry Based Untargeted Metabolomics.

Guo Jian J   Huan Tao T  

Analytical chemistry 20200527 12


Full-scan, data-dependent acquisition (DDA), and data-independent acquisition (DIA) are the three common data acquisition modes in high resolution mass spectrometry-based untargeted metabolomics. It is an important yet underrated research topic on which acquisition mode is more suitable for a given untargeted metabolomics application. In this work, we compared the three data acquisition techniques using a standard mixture of 134 endogenous metabolites and a human urine sample. Both hydrophilic i  ...[more]

Similar Datasets

2020-07-14 | MSV000085744 | GNPS
2020-10-02 | MTBLS1842 | MetaboLights
2018-03-26 | MTBLS417 | MetaboLights
2017-11-20 | PXD006934 | Pride
2023-12-19 | PXD045318 | Pride
| MSV000095787 | GNPS
2023-11-21 | PXD041421 | Pride
2023-11-21 | PXD041391 | Pride
2020-02-17 | PXD016662 | Pride
2023-07-11 | PXD037527 | Pride